Chapter 2.13
Tousimis 815 Critical-Point Dryer (small samples)
(cpd2)
1.0 Title
Tousimis Model 815 Autosamdri CO2
critical-point drying system
2.0 Purpose
This document has specific information about the
capabilities, configuration, and proper operation of the Model 815
critical-point drying system.
3.0 Scope
3.1
The Autosamdri-815 (cpd2) is a fully automatic CO2
critical point drying apparatus. Its front panel consists of a pressure gauge,
a temperature gauge, a purge timer knob, and six process push buttons (COOL,
FILL, PURGE, HEAT, BLEED, and VENT) with green LEDs. Only COOL and FILL buttons
need to be pushed for a complete process. The unit can also operates under
semi-automatic mode, but it is not recommended for the inexperienced users. The
sample holders for cpd2 are 1-inch in diameter.
3.2
On the flat panel of the unit, there are two handle screws,
that control the FILL and PURGE/VENT flow. They are factory set. DO NOT ADJUST
THE VALVES.
3.3
If you feel that the
PURGE flow rates have been altered; there is an easy to adjust via a 'Dry Run'
(that is without any alcohol or sample. This way you can visualize the flow of
the LCO2 through the chamber via the chamber window:
3.3.1
Make sure that both FILL and PURGE meter valves are free to
turn by making sure that both 1/32" hex head screws are backed off and
loose so you may freely turn the valves.
3.3.2
Turn CPD on.
3.3.3
Set Purge Timer to pos# 2 (10 min.) to allow for enough time
to comfortably make the following adjustments.
3.3.4
Press COOL.
3.3.5
Press FILL after COOL shut off so that you may visualize the
incoming LCO2 from the left hand side of the chamber. The LCO2 will look like a
steady stream entering the chamber. As the chamber level fills with LCO2; the
stream will disappear.
3.3.6
After the 2 minute FILL mode; the process will advance into
the PURGE mode. This is where the most critical adjustment is necessary.
[It is essential that the incoming FILL rate exceeds the
outgoing PURGE rate so that the LCO2 level does not drip below the level of the
devices being dried. You also; want to be sure that you have steady flow so
that exchange of alcohol for LCO2 occurs. Therefore; the PURGE meter valve is
adjusted one way another until you see movement from left to right within the
chamber (indicating flow) until the 1st hint of small bubbles are evident. Stop
at this point. This is the point where you have good flow; yet the LCO2 level
is not dropping.]
3.3.7 Let the 815 process complete its course
automatically.
4.0
Applicable Documents
Pressure vs.
Temperature Graph for CPD Process
Verified
Temperature Pressure Graph of Microlab CPD Process
5.0
Definitions, Process Terminology,
Nomenclature
5.1
CPD: critical-point drying (dryer)
5.2
LCO2: liquid carbon dioxide
6.0 Safety
6.1
The CPD system normally operates under very high chamber
pressures. DO NOT DEVIATE FROM THE CORRECT PROCEDURE FOR LOADING THE SAMPLE AND
SECURING THE CHAMBER LID – personal injury is quite possible otherwise.
6.2
An uncontrolled release of CO2 can cause personal
injury due to frostbite. Again, DO NOT DEVIATE from the correct procedures.
Make very sure that persons near the system are always wearing protective eye
ware.
7.0 Statistical/Process
Data
8.0 Available
Processes, Process Notes
8.1
CPD uses copious amounts of LCO2. Requests for
LCO2 cylinder changes are the primary CPD problem report and result
in decreased tool availability for all lab members. Review the following
Process Notes and attempt to minimize your LCO2 usage.
8.2
Extra long purge time does NOT increase the effectiveness of
your critical point drying process; it just wastes LCO2.
8.3
Minimize the amount of alcohol you add to the chamber. You
only need enough alcohol to just cover your parts.
8.4
Maximize the volume of the CPD chamber filled with Teflon® inserts. The more Teflon® in the chamber, the less alcohol in the
chamber, and the less LCO2 needed.
Most of the LCO2 is consumed during the PURGE process. Each
number of the PURGE dial indicator corresponds to 5 minutes of flowing LCO2;
i.e., setting 1 = 5 min, setting 2 = 10 min, setting 4 = 20 min, etc. The goal
of the purge process is to replace essentially all the methanol/IPA in the
chamber with LCO2. Since methanol and IPA are especially soluble in
LCO2, there is NOT a linear relationship between purge time and
alcohol removal. Most alcohol is removed from the chamber after only one LCO2
volume exchange. Once the alcohol has been removed, the LCO2 can be
heated to supercritical conditions and a phase change from liquid to vapor can
take place without crossing a liquid/vapor equilibrium line and generating the
surface tension forces that result in adhesion of released MEMS parts. (See
Section 4, Pressure
vs. Temperature Graph for CPD Process).
Process staff has verified that there was no remaining alcohol and therefore a completely effective CPD process was performed with the following samples and volumes of alcohol.
Sample Holder |
Teflon® |
Steel |
None |
Amt Methanol |
10 ml |
17 ml |
25 ml |
Purge Setting |
1 |
1 |
2 |
Purge Time |
6.5 min |
6.9 min |
11.17 min |
Cool CO2 Usage |
0.5 lb |
0.5 lb |
0.5 lb |
Chamber CO2 Usage |
1.5 lb |
1.0 lb |
3.0 lb |
A purge setting of 1 is sufficient for most needs and there should not be any need for a purge setting greater than 2.CPD is often one of the last steps of a lengthy MEMS process. Users are wary to risk their parts to an ineffective CPD process. If you would like to verify that a lower purge time is suitable for your process – run a dummy process one time. Add a non-critical sample and just enough alcohol to cover the sample. Use the recommended minimum purge time. Verify that the chamber is absolutely dry (no residual alcohol) at the end of the process. If the chamber is dry, you have performed a fully effective CPD process.
If you ever observe residual alcohol in the
chamber using the minimum recommended times, report it as a CPD problem on the
wand.
9.0 Operation
9.1
Preparation
9.1.1
Before starting to use cpd2, the user should check the
liquid CO2 cylinder weight. If the weight is 25 lbs (or more) below
its full weight, report a fault on the WAND. This will give the staff adequate
time to change the tank. Typically, each run consumes approximately 1 lb and a
full tank provides 30 lbs of CO2 according to the vendor. Therefore,
the cylinder has to be changed before 30 lb has been used up.
9.1.2
Clean samples carefully. Photoresist residue may re-deposit
on the dried samples or the chamber walls.
9.1.3
The user should also decide on the purge time beforehand and
set the purge timer knob accordingly. A setting of 2 indicates 10 minutes, a
setting of 3 indicates 15 minutes, and so on to a maximum of 30 minutes. See Section 8 for Process Notes
regarding appropriate Purge times.
9.2
Operation
9.2.1
Check for reservations on the WAND for cpd or cpd2. The two
systems share one CO2 tank, therefore they cannot be used
simultaneously. Likewise, if either of the systems is enabled already, the ON
indicator above the system (on the front of the hood) will be illuminated.
9.2.2
Enable cpd2 on the WAND. Note the ON indicator at the
front of the hood lights up.
9.2.3
Ensure that the three-way manual valve on the front of the
hood points toward CPD2. If not, rotate the handle 180° until it does.
9.2.4
Verify that the cylinder valves of both liquid CO2
cylinders are open. Check the weight of the cylinders. Both liquid CO2 cylinders
are located behind the tools on the other side of the 432B wall. One is labeled
CHAMBER CO2 and the other, COOL CO2. Report a
fault on the WAND if the weight is 20 lbs or more below its full weight.
9.2.5
Record the weight of both CO2 cylinders (see
9.2.18 below).
9.2.6
Clean the chamber and lid using methanol/IPA and dry with N2
gun. Repeat several times if necessary. Use Techni-cloth cleanroom wipes.
9.2.7
Let unit warm up for 3-5 minutes. This will stabilize all
plumbing components, the metering valves, and the Static Pressure Control
module.
9.2.8
Press the COOL button. The LED on the button will go on
(VENT LED will go off). Now place only enough Methanol or IPA into the chamber
to cover your samples.
9.2.9
Place your sample(s) in the chamber. There are three steel
sample holders available, and one Teflon® sample holder for 10mm samples. The sample holders are stored in
a drawer below the cpd systems. Alternatively, some users have custom sample
holders, and in fact the system can be used without any sample holder at all,
by placing the sample on the bottom of the chamber.
Be careful
to keep the surface of the sample wet while transferring it into the chamber.
Practice and experience are very helpful here.
9.2.10
Place the lid on the chamber. The side with the tousimis
stamp should face up. Make sure that the O-ring is in place. Finger-tighten three knurled nuts evenly. DO
NOT USE A WRENCH OR OTHER TOOLS. The unit will go on cooling by itself until
slightly below 0°C. If it takes more than 2 minutes to reach 0°C, the liquid CO2
cylinder may be depleted.
9.2.11
Once the cooling reaches ~0°C, press the FILL button. The
unit will start filling the chamber with liquid CO2. The user can
see a visible liquid flow through the fill port. If not, the liquid CO2 cylinder may be depleted. Now
the unit is in the AUTOMATIC mode. It is not necessary to push any more buttons
till the process has completed.
9.2.12
After 2 minutes, the unit switches to PURGE Mode automatically.
The PURGE LED will be on. The Methanol/IPA is exchanged with liquid CO2. The PURGE time depends on the setting of the
timer knob the user set before the process starts. See Section 8 for important notes on
PURGE time settings. There should be no liquid coming out the chamber at
the end of PURGE time. The unit will switch to the HEAT mode. The HEAT LED will
be on.
9.2.13
When the pressure reaches 1250 psi and the temperature is
above 31°C, the Tousimis Equilibrium cycle starts. At this point the HEAT LED
will be blinking for 4 minutes.
9.2.14
At the end of the Tousimis Equilibrium, the unit switches to
BLEED cycle and the BLEED LED will be on. The Bleed flow should be around 7
SCFH for pressures between 1,200 and 400 psi.
The temperature should be above 31°C.
9.2.15
There may be a hissing or cracking sound in the BLEED
cycle. It is normal because gaseous CO2
or dry ice may be coming out the unit.
9.2.16
When the pressure drops below 360 psi, the unit switches to
VENT mode. VENT LED will be on. Do not try to loosen the nuts until the chamber
is completely vented. It will damage the threads on the nuts and the chamber.
After 3 to 4 minutes, the chamber will come to the atmospheric pressure. NOW the user should be able to loosen the
knurled nuts with fingers. It is time to open the chamber and remove your
samples.
9.2.17
Clean and blow-dry the chamber and lid with methanol/IPA if
you see any residue. Close the lid to keep the chamber clean and moisture free.
9.2.18
Leave the cylinder valves of both liquid CO2
cylinders open. Report the initial and final cylinder weights on the WAND
comment when disabling cpd2. If either of the final cylinder weights is below
20 lb of the full weight, report fault on the WAND also.
10.0
Troubleshooting Guidelines
For
most structures, there should be no reason to adjust the Fill or Bleed needle
valves. These valves are adjusted and set - so that there is slow flow through
the chamber during the purge process - but the LCO2 level stays constant - and
the chamber remains almost completely full of LCO2.
The most convenient way to verify that these valves are properly set is to allow a minimum head space in the chamber just below the viewing window. The minimum head space enables viewing of just a few mm size bubbles through the chamber window. Track these bubbles by eye during the purge process so that the flow through the chamber enables a 'bubble travel speed' of about 1inch/second.
The small cpd tool (cpd2) has small allen set screws to lock the needle valves in position. Once the valves have been properly set, staff use the small allen set screws to lock the needle valves in position. Though there is a chamber inlet and exit filter, these valves sometimes become partially blocked and the needle valves need to be readjusted. For cpd2, if you need to readjust the needle valves, be sure to loosen the allen set screws prior to attempting to turn the needle valve. Otherwise you will strip the needle valve handle. Then follow the procedure above to set the needle valves and re tighten the needle valve set screws. The large cpd tool (cpd) does not have locking set screws and the needle valves may be adjusted without ‘unlocking’ if needed.
Even though the fill needle valve
prevents a sudden in-rush of LCO2, some members report that the large pressure
change when the solenoid opens can damage structures with a large surface area
compared to the flexures that support them. (e.g. mirrors!) To prevent
11.0
Qualification Procedures
11.1
Potential new users should arrange ahead of time to
have an existing user review cpd2 operation. Casual questioning (e.g.
"Hey, can you review this with me for a moment?") is not to be
considered an official review period.
11.2
The first question that should be asked is whether Section
2.7, “Tousimis Critical Point Dryer" of the WAND has been thoroughly
reviewed? If not, insist that it be
reviewed before the meeting.
11.3
The "meeting" between the current and new user
should be scheduled for enough time to explain equipment operation, load &
unload sequence, etc.
Equipment
time can be recharged to new user’s account if necessary.
11.4 Step 11.3 should
be done at least twice by two different current users.
11.5
Superusers will first ask, “who trained you?” of the new
user. “Training” implies adequately following Steps 11.1-4 above. The users
identified as being the ones who did the training will be responsible for the
information conveyed by the new user during the qualification.
Teflon®
is a registered trademark of DuPont.