Chapter
2.14
Tousimis 915B Critical-Point
Dryer
(cpd)
1.0
Title
Tousimis Model 915B Automegasamdri® CO2
critical-point drying system

Figure 1 -
Tousimis 915B Critical Point Drying System
2.0
Purpose
This document has specific information about the capabilities,
configuration, and proper operation of the Model 915B critical-point drying
system.
3.0
Scope
3.1
The Automegasamdri-915B (cpd) is a
fully automatic CO2 critical point drying apparatus (Figure 1). Its
front panel consists of a pressure gauge, a temperature gauge, a purge timer
knob, and six process push buttons (COOL, FILL, PURGE, HEAT, BLEED, and VENT)
with green LEDs. Only COOL and FILL buttons need to be pushed for a complete
process. The unit can also operate under semi-automatic mode, but it is not
recommended for the inexperienced users. The sample holders for cpd are
available for 2-inch through 6-inch wafers. Holders for the Tousimis 815 (cpd2)
can also be used in cpd
3.2
On the flat panel of the unit,
there are four micrometer valves that control the COOL, BLEED, FILL and
PURGE/VENT flows. They are set during the install, and should not need
readjustment. DO NOT ADJUST THE VALVES.
4.0
Applicable
Documents
4.1
Automegasamdri-915B User Manual,
Serial Number 1257, Tousimis Inc.
4.2
Tousimis
915B Operating Manual (24 MB PDF file)
4.3
Pressure vs.
Temperature Graph for CPD Process
4.4
Verified
Temperature Pressure Graph of Microlab CPD Process
4.5
Cpd-Plumbing Schematics
by Mike Linan (11/7/07)
5.0
Definitions,
Process Terminology, Nomenclature
5.1
CPD: critical-point drying (dryer)
5.2
LCO2: liquid carbon
dioxide
6.0
Safety
6.1
The CPD system normally operates
under very high chamber pressures. DO NOT DEVIATE FROM THE CORRECT PROCEDURE
FOR LOADING THE SAMPLE AND SECURING THE CHAMBER LID – personal injury is quite
possible otherwise.
6.2
An uncontrolled release of CO2
can cause personal injury due to frostbite. Again, DO NOT DEVIATE from the
correct procedures. Make very sure that persons near the system are always
wearing protective eyeware.
7.0
Statistical/Process
Data
N/A
8.0
Available
Processes, Process Notes
8.1
CPD uses copious amounts of LCO2.
Requests for LCO2 cylinder changes are the primary CPD problem
report and result in decreased tool availability for all lab members. Review
the following Process Notes and attempt to minimize your LCO2 usage.
8.2
Extra long purge time does NOT
increase the effectiveness of your critical point drying process; it just
wastes LCO2.
8.3
Minimize the amount of alcohol you
add to the chamber. You only need enough alcohol to just cover your parts.
8.4
Maximize the volume of the CPD
chamber filled with Teflon® inserts. The more Teflon® in
the chamber, the less alcohol in the chamber, and the less LCO2
needed.
8.5
Most of the LCO2 is consumed
during the PURGE process. Each number of the PURGE dial indicator corresponds
to 5 minutes of flowing LCO2; i.e., setting 1 = 5 min, Setting 2 = 10 min,
Setting 4 = 20 min, etc. The goal of the purge process is to replace
essentially all the methanol/IPA in the chamber with LCO2. Since
methanol and IPA are especially soluble in LCO2, there is NOT a
linear relationship between purge time and alcohol removal. Most alcohol is
removed from the chamber after only one LCO2 volume exchange. Once
the alcohol has been removed, the LCO2 can be heated to
supercritical conditions and a phase change from liquid to vapor can take place
without crossing a liquid/vapor equilibrium line and generating the surface
tension forces that result in adhesion of released MEMS parts. (See Section 4,
Pressure vs. Temperature Graph for CPD Process).
8.6
Process staff has verified that
there was no remaining alcohol and therefore a completely effective CPD process
was performed with the following samples and volumes of alcohol.
|
4” Wafer |
||
|
Number of wafers Amount of Methanol Purge Setting Cool CO2 Chamber CO2 |
4 120 ml 4 4.5 lb 10.5 lb |
1 40 ml 2 5.0 lb 3.5 lb |
|
6” Wafer |
||
|
Number of wafers Amount of Methanol Purge Setting Cool CO2 Chamber CO2 |
4 320 ml 6 5.5 lb 13.5 lb |
1 80 ml 2 4.5 lb 7.5 lb |
8.7
For a single 4” or 6” wafer there
is never a need for a purge setting greater than 2.
8.8
CPD is often one of the last steps
of a lengthy MEMS process. Users are wary to risk their parts to an ineffective
CPD process. If you would like to verify that a lower purge time is suitable
for your process – run a dummy process one time. Add a non-critical sample and
just enough alcohol to cover the sample. Use the recommended minimum purge
time. Verify that the chamber is absolutely dry (no residual alcohol) at the
end of the process. If the chamber is dry, you have performed a fully effective
CPD process.
8.9 If you ever observe residual alcohol in the chamber using the minimum recommended times, report it as a CPD problem on the wand.
9.0
Operation
9.1
Preparation
9.1.1
Before starting to use cpd, the
user should check the liquid CO2 cylinder weights. See Figure 2. If
either of the weights is 25 lbs (or more) below their full weights, report a
fault on the WAND. This will give the staff adequate time to change the
tank(s). A full tank provides 30 lbs of CO2 according to the vendor.
Therefore, the cylinder has to be changed before 30 lb has been used up.

Figure
2 - CPD and CPD2 Cylinder Scale Readouts
9.1.2
Clean samples carefully.
Photoresist residue may re-deposit on the dried samples or the chamber walls.
9.1.3
The user should also decide on the
purge time beforehand and set the purge timer knob accordingly. A setting of 1
indicates 5 minutes; a setting of 2 indicates 10 minutes; and so on, to a
maximum of 45 minutes. See Section 8 for Process Notes regarding
appropriate Purge times.
9.2
Operation
9.2.1
Check for reservations on the WAND
for cpd or cpd2. The two systems share one CO2 tank;
therefore they cannot be used simultaneously. Likewise, if either of the
systems is enabled already, the ON indicator above the system (on the
front of the hood) will be illuminated.
9.2.2
Enable cpd on the WAND. Note the ON
indicator at the front of the hood lights up (Figure 3). The green LED on the
VENT button (on the cpd console) will go on. This means the unit is in the vent
mode.

Figure
3 - CPD Enable Indicator
9.2.3
Ensure that the three-way manual
valve (Figure 4) on the front of the hood points toward CPD. If not,
rotate the handle 180° until it does.

Figure 4 - CO2
Selection Valve
9.2.4
Verify that the cylinder valves of
both liquid CO2 cylinders are open. Check the weights of the cylinders. Both liquid CO2 cylinders
are located behind the tools on the other side of the 432B wall. One is labeled
CHAMBER CO2 and the other, COOL CO2. Report a
fault on the WAND if the weight is 20 lbs or more below its full weight.
9.2.5
Record the weight of both CO2
cylinders (see 9.2.25 below).
9.2.6
Remove the eight steel nuts and
eight bronze washers from the chamber lid. Place a clean Techni-cloth or two on
the table top. Using the index fingers of both hands, insert your fingers into
the loops on the lid and lift the lid from the chamber and place it onto the
Techni-cloths. The lid is heavy – do this slowly and carefully. Do not allow
the lid bottom to contact the top of the steel bolts – the bolts will scratch
the lid sealing surface and cause it to leak.
9.2.7
Remove the
four Teflon® spacer rings from the chamber (Figure 5). Remember to wear poly gloves (clear) when you are handling Teflon®
spacer rings.

Figure 5 - Teflon®
Spacer Rings (4)
9.2.8
Clean the chamber using
methanol/IPA and dry with N2 gun. Repeat several times if necessary.
Do not use the handi-wipes from the 432B hood – they shed lint, and the
lint will end up all over your sample as well as the chamber. This has been
experimentally verified. Use only Techni-cloth cleanroom wipes. Refit the appropriate Teflon®
spacer rings back into the chamber:
|
Sample |
Rings to Use |
|
6” wafer |
None |
|
4”
wafer |
One
(the largest) |
|
3”
wafer |
Two
largest |
|
2”
wafer |
Three
largest |
|
Something
smaller |
All
four |
9.2.9 As you replace the rings into the chamber, wipe
them down with a Techni-cloth and IPA/methanol. Note that when all four rings
are in the chamber, the available sample size is the same as for cpd2.
9.2.10
Place only
enough Methanol or IPA into the chamber to cover your sample(s). More alcohol
requires a longer purge time (see 9.1.3), and thus more CO2.
9.2.11
Place your sample(s) in the
chamber. There are five Tefflon®
sample holders available: 6”, 4”, 3”, 2”, and pieces-parts basket. The
sample holders are stored in a drawer below the cpd systems. Alternatively, some
users have custom sample holders, and in fact the system can be used without
any sample holder at all, by placing the sample on the bottom of the chamber.
Sample holders from cpd2 can also be used.
Be careful to keep the surface of
the sample wet while transferring it into the chamber. Practice and experience
are very helpful here.
9.2.12
Place the lid on the chamber. Take
great care to avoid allowing the bottom of the lid to make contact with the top
of the steel studs. Once the bolt holes in the lid line up with the studs, as
viewed from directly above, carefully lower the lid into place.
9.2.13
Press the COOL button. The LED on
the button will go on (VENT LED will go off). The unit will go on cooling by
itself until slightly below 0°C. If it takes more than 2 minutes to reach 0°C,
the COOL LCO2 cylinder may be depleted.
Note that we could easily have
begun the cool down cycle before placing the lid on the chamber. Why might this
be a bad thing? If we were to cool down the chamber before installing the lid,
we run the risk of condensing moisture from the lab air. By cooling the chamber
after placing the lid in place, we minimize that risk. The cool down
takes only a couple of minutes anyway, so time savings is not an issue.
9.2.14 Replace the eight bronze flat washers, then
replace the eight steel nuts, but do not tighten past gentle contact.
9.2.15 Using the spanner wrench, and working in a star
crossing pattern, begin tightening the nuts one at a time. It is critically
important to tighten the bolts evenly to avoid leakage and damage to the lid or
the chamber. See Figure 6. Make two passes around the lid in the star pattern. Proper
tightening does not require a lot of force – don’t try to kill it!

Figure 6 -
Lid Tightening Sequence
9.2.16
Verify that the chamber
temperature has reached 0° C, then press the FILL button. The unit will start
filling the chamber with liquid CO2. The user can see a visible
liquid flow through the fill port. If
not, the CHAMBER LCO2 cylinder may be depleted. Now the unit
is in the AUTOMATIC mode. It is not necessary to push any more buttons until
the process has completed.
9.2.17
After 8 minutes, the unit switches
to PURGE Mode automatically. The PURGE LED will be on. The Methanol/IPA is
exchanged with liquid CO2. The PURGE time depends on the setting of
the timer knob the user set before the process starts. See Section
8 for important notes on PURGE time settings.
9.2.18
Upon Completion at the PURGE mode,
the unit will advance into a POST-PURGE-FILL mode in which the chamber fills
with LCO2 for an additional 4 minutes. This will be indicated by
illumination of both the FILL and PURGE LEDs.
9.2.19
The unit will then switch to the
HEAT mode. The HEAT LED will be on.
9.2.20
When the pressure reaches around
1350 psi and the temperature is above 31°C, the Tousimis Equilibrium cycle
starts. At this point the HEAT LED will be blinking for 4 minutes.
9.2.21
At the end of the Tousimis
Equilibrium, the unit switches to BLEED cycle and the BLEED LED will be on. The
Bleed flow should be around 40-50 SCFH for pressures between 1,300 and 400 psi.
This setting should yield an average of approximately 100-150 psi/min reduction
in pressure. The temperature should be above 31°C.
9.2.22
There may be a hissing or cracking
sound in the BLEED cycle. It is normal
because gaseous CO2 or dry ice may be coming out the unit.
9.2.23
When the pressure drops below 300
psi, the unit switches to VENT mode. The VENT LED will be on. Do not try to
loosen the nuts until the chamber is completely vented - it will damage
the threads on the nuts and the chamber. After 3 to 4 minutes, the chamber will
come to atmospheric pressure. NOW the
chamber lid may be removed by alternatively and evenly loosening all of the steel
nuts using the spanning wrench. Remove the nuts, the bronze washers, the lid (carefully)
and the (dry) sample.
9.2.24
Clean and blow dry the chamber and
lid with methanol/IPA if you see any residue. Replace ALL of the Teflon® spacer
rings. Close the lid to keep the chamber clean and moisture free. Replace the
flat washers and the steel nuts, but do not tighten them.
9.2.25
Leave the cylinder valves of both
liquid CO2 cylinders open. Report the initial and final cylinder
weights on the WAND comment when disabling cpd. If either of the final cylinder
weights is below 20 lb of the full weight, report a fault on the WAND also.
10.0
Troubleshooting
Guidelines
10.1 A backing ring has been added to the chamber o-ring groove which is supposed to extend the life of the Teflon o-ring greatly. The backing ring looks like a thin, somewhat flat o-ring with a triangular cross-section. It is located under the normal o-ring and should not be removed during o-ring changes. If it comes out with the o-ring, it should be re-installed just as it came out. Care should be taken to ensure this as it can be installed upside down. A spare o-ring is in the cpd drawer beneath the tool. More spares are available from the stockroom.
10.2 There should be no reason to adjust the Fill or Bleed needle valves. These valves have been adjusted and set - so that there is slow flow through the chamber during the purge process - but the LCO2 level stays constant - and the chamber remains almost completely full of LCO2.
The most convenient way to verify that these valves are properly set is to allow a minimum head space in the chamber just below the viewing window. The minimum head space enables viewing of just a few mm size bubbles through the chamber window. Track these bubbles by eye during the purge process so that the flow through the chamber enables a 'bubble travel speed of about 1inch/second.
Once
the valves have been properly set, staff use the small allen set screws to lock
the needle valves in position. Though there is a chamber inlet and exit filter,
these valves sometimes become partially blocked and the needle valves need to
be readjusted. If you need to readjust the needle valves, be sure to loosen the
allen set screws prior to attempting to turn the needle valve. Otherwise you
will strip the needle valve handle. Then follow the procedure above to set the
needle valves and re-tighten the needle valve set screws.
10.3 If you enable cpd and find the vent LED not lit on, check the main power of cpd. It might be OFF by accident.
11.0
Qualification
Procedures
11.1
Potential new users should arrange
ahead of time to have an existing user review cpd operation. Casual
questioning (e.g. "Hey, can you review this with me for a moment?")
is not to be considered an official review period.
11.2
The first question that should be
asked is whether Section 2.8, “Tousimis Critical Point Dryer" of the WAND
has been thoroughly reviewed? If not,
insist that it be reviewed before the meeting.
11.3
The "meeting" between
the current and new user should be scheduled for enough time to explain
equipment operation, load & unload sequence, etc. Equipment time can be
recharged to new user’s account if necessary.
11.4
Step 3 should be done at least
twice by two different current users.
11.5
Superusers will first ask, “who
trained you?” of the new user. “Training” implies adequately following Steps
11.1-11.4 above. The users identified as being the ones who did the training
will be responsible for the information conveyed by the new user during the
qualification.
Teflon®
is a registered trademark of DuPont.