Chapter 3.4
APT Emulsion Process
(aptemul)
1.0
Title
APT Emulsion Process
2.0
Purpose
The APT Emulsion Developer
is an automated photomask processing unit for emulsion masks which have been exposed in the pattern generator.
3.0
Scope
This chapter describes the
emulsion process used in the APT emulsion unit on the right. The programs to
which we refer in this manual are listed in Section 9.3.
Please note that sink12 must be enabled when using the APT. The aptemul mask
developer accommodates masks in the following sizes: 2.5”, 4”, 5”, 6” and 7”.
4.0
Applicable Documents
5.0
Definitions and Process Terminology
6.0
Safety
Before you proceed, please note the following: the aptemul chemical
tubings should always be kept immersed in solution to avoid spraying air onto
the plate instead of the desired chemical. The door to the process chamber
should be kept closed and latched at all times except when loading or unloading
masks.
7.0
Statistical/Process Data
8.0
Available Process, Gases, Process Notes
9.0
Equipment Operation
►
Enable the aptemul on the Wand. Press the round green power button
to turn on the system power.
►
Open front panel.
►
Put the colored tubings from the beakers into the required
premixed chemical bottles and cap them tightly. Refer to the Program listings
in Section 9.3 to determine which chemicals you will
need, and to the table below to see the tubing assignments for each chemical.
|
Tubing Color |
Chemical |
|
Green |
PD-86 (premixed) Negative Developer |
|
Orange |
Kodak Rapid Fixer |
|
Blue |
Methanol |
|
White |
D.I. Water (main supply) |
Developers age when exposed to light or air, so cap tightly and
keep the front panel of the aptemul closed. If the PD-86 developer appears to
be discolored, discard whatever is left in the bottle and replace it with a
fresh bottle (Cabinet #3, in the lobby, on the bottom shelf.)
9.1
Checking Chemical Lines
Before starting your process, you will run a test program (Program 82 - Negative Process Test) to check that each
chemical is flowing properly at each step. These programs are set up to allow
each required chemical to spray in sequence, with water rinses in between. Program 82 is listed in Section 9.3.
9.1.1
Enter test Program 82.
9.1.2
Press START/RESUME button.
9.1.3
When the program starts, make sure that the desired chemical
starts to spray out of its respective nozzle. The chemical(s) should dispense
from the nozzle(s) as a fan-shaped spray. When it has been spraying for 30
seconds, press SKIP button. This will advance the program to its next
step, which is a DI water rinse.
9.1.4
Let the water rinse cycle run through (30 seconds). The next
chemical should then begin to spray. Once it has been spraying 30 seconds,
press SKIP again to start a water rinse. Continue in this way until you have
worked your way through the program. If too much water beads up on the window
and prevents you from being able to see through, press the HOLD button
near the end of a water cycle. Open the window and wipe it dry with a
Techni-cloth to clear it. Techni-cloths are kept in a dispenser on top of the
aptchrome. Press the START/RESUME button to continue the cycle. To go
back to a given step, press the ABORT button and then the SKIP
button until you reach the desired step, then press START/RESUME. Continue
in this way until you have checked the process parameters of the entire
program. The lamp is on all the time on the test programs and the pump rinse
programs.
9.1.5
If everything checks out, proceed to Section
9.2. If there is a problem, ask the process staff for help and post the
problem on Faults. Note: There
is a new feature at the aptemulsion: a holding tank for the rapid fixer
effluent. EH&S will no longer permit the Microlab to have the Kodak Rapid
Fixer go down the sink drain after the fix step (Programs
82 and 2). In the service chase CR1, behind the
APTs, there is now a holding tank for this waste. When the holding tank is
full, two small orange warning lights (indicated at Input1 and Input 2A) will
go on. Until this tank is drained (by the aptemul technician), NO RAPID
FIXER IS DISPENSED THROUGH THE LINE. Although the tank level will be
checked weekly by the process staff, it will be required by each qualified user
to check for the warning lights before enabling and operating this piece of
equipment.
9.2
Developing Masks: Red Lights ON in R1 and R2!
Note: Before developing your first plate, run Program 2 through 1X to insure all lines are clear.
9.2.1
Load mask, emulsion side up, onto the spinner, seating all 4
corners securely so that it does not fly off and break during spin.
9.2.2
Close and latch sliding window and enter Program 2 - Negative
Process.
9.2.3
Press START.
9.2.4
When the process is finished, check to see if the plate is
completely dry. If not, blow it dry with the antistatic N2 gun in R2 mounted at
the pattern generator chamber.
9.2.5
When you have finished processing your masks, open the front
panel. Check whether the beakers are filled to the top with DI water. The deck
hose at sink12 has a curly cord long enough to reach the aptemul. Please refill
when the beakers are less than half full.
9.2.6
Put tubing back into the designated beakers and recap the chemical
bottles.
9.2.7
Run Program 92 to rinse out the process
pumps and prevent damage to them.
9.2.8
When the entire cycle for Program 92 is finished, open the
window, use the deck hose at sink12 to rinse the chamber.
9.2.9
Close and latch the window and run Program 93
to water rinse it for 30 more seconds.
9.2.10
Use a techni-cloth to wipe the interior of the window. Disable the
aptemul on the Wand.
IMPORTANT: Rinse out the methanol pump immediately with water after use, or
the pump will fail after the vapor pressure of the methanol builds up and
swelling occurs inside the pump.
Never change programs that are listed on the instructions sheets
or the programs, which are locked.
Program 82 – Negative
Process Test
|
Step |
Chemical |
Output |
Description |
Time (sec.) |
|
1 |
PD-86 (premixed) |
4 |
Developer 1 |
100 |
|
2 |
Water Rinse |
7 |
Rinse Nozzle |
30 |
|
3 |
Kodak Rapid Fixer |
2 |
Fixer |
100 |
|
4 |
Water Rinse |
7 |
Rinse Nozzle |
30 |
|
5 |
Methanol |
5 |
Methanol |
100 |
|
6 |
Water Rinse |
7 |
Rinse Nozzle |
30 |
Program 2 – Negative
Process
|
Step |
Chemical |
Output |
Description |
Time (sec.) |
|
1 |
Water Rinse |
7 |
Rinse Nozzle |
10 |
|
2 |
PD-86 (premixed) |
4 |
Developer 1 |
80 |
|
3 |
Water Rinse |
7 |
Rinse Nozzle |
45 |
|
4 |
Kodak Rapid Fixer |
2 |
Fixer |
60 |
|
5 |
Water Rinse |
7 |
Rinse Nozzle |
45 |
|
6 |
Water, Methanol |
5 or 7 |
Methanol/Rinse Nozzle |
30 |
|
7 |
Methanol |
5 |
Methanol |
30 |
|
8 |
Spin Dry |
- |
Spindry |
120 |
Program 92 – Negative
Process Pump Rinse
|
Step |
Chemical |
Output |
Description |
Time (sec.) |
|
1 |
PD-86 (premixed) |
4 |
Developer 1 |
30 |
|
2 |
Water Rinse |
7 |
Rinse Nozzle |
30 |
|
3 |
Kodak Rapid Fixer |
2 |
Fixer |
30 |
|
4 |
Water Rinse |
7 |
Rinse Nozzle |
30 |
|
5 |
Methanol |
5 |
Methanol |
30 |
|
6 |
Water Rinse |
7 |
Rinse Nozzle |
30 |
Program 93 – Water Rinse
|
Step |
Chemical |
Output |
Description |
Spin Speed (103
RPM) |
Time (sec.) |
|
1 |
DI Water Rinse |
7 |
Developer 1 |
.22 |
30 |
|
2 |
Spin Dry |
Lamp |
Spindry |
.95 |
120 |
10.0
Troubleshooting Guidelines
11.0
Figures and Schematics
12.0
Appendices
12.1
How To Make Up a New Batch of Rapid Fixer
12.1.1
First, take the nearly empty bottle of rapid fixer out of the
aptemul chamber and over to the sink 12 deck (wear appropriate safety gear).
12.1.2
Next, discard the remaining fixer at sink 12 and rinse out the
bottle.
12.1.3
The second C-locker (R1 C-locker #2) in the red room is where
another package of Kodak Rapid Fixer can be found; if not, check the blue
developer cabinet - it's in a brown cardboard box on the bottom shelf.
12.1.4
Fill the now empty glass 1-gallon bottle halfway with DI water,
then add Solution "A" to the bottle. Next, add Solution "B"
(hardener). Recap the bottle tightly and agitate the solutions so that they mix
well.
12.1.5
Finish up by adding more DI water to the shoulder of the bottle.
It will be slightly foamy, so add the water slowly. Return the bottle to the
aptemul chamber.
Note: The
solution MUST be covering the tubing in the bottle at all times; when it
does not, it's time to make up a new batch.
12.2
Processing Menu
