Chapter 3.7
Ultratech Mask Copier
(maskcopy)
1.0 Title
Ultratech Mask Copier
2.0 Purpose
The Ultratech mask copier is a contact printer, which is used to
duplicate both chrome and emulsion masks.
3.0 Scope
This document
describes the general operation of the Ultratech mask copier. It is a contact
printer primarily used for duplicating chrome and emulsion masks. The chamber
door has to be changed to accommodate various size plates. Currently the lab
has fixtures to copy 2.5 and 5 square plates. The copy will be a mirror image
of the original mask. The original mask is called the master plate.
4.0 Applicable
Documents
5.0 Definitions &
Process Terminology
5.1 Master
plate: The original mask.
5.2 Copy
plate: The duplicate mask.
6.0 Safety
Follow general safety guidelines in the lab as well as the
specific safety rules as per follows:
6.1 Take care of the vacuum seal rubber. Do not cut the rubber ring with the edges of the plates during loading and unloading; this will cause a vacuum leak.
7.0 Statistical/Process
Data
Pertinent
information can be found in the following locations:
7.1 Problem and comments section under
equipment section of the wand.
7.2 Enable message for maskcopy.
8.0 Available
Processes, Gases, Process Notes
8.1 Available
Processes
8.1.1 Exposure time and
intensity settings for 2.5 plate
|
Plate Type |
Emulsion Intensity Setting |
Chrome Intensity Setting |
Exposure Time (sec.) |
|
P.R. on Chrome |
0 |
3.5 |
60.0 |
8.1.2 Exposure time and
intensity settings for 5 plates
|
Plate Type |
Emulsion Intensity Setting |
Chrome Intensity Setting |
Exposure Time (sec.) |
|
P.R. on Chrome |
0 |
3.5 |
60.0 |
|
Emulsion |
3.0 |
0 |
3.0 |
Note: Only ~ 4.5 of the exposing field
is guaranteed to be uniform.
9.0 Operating
Procedure
The following is a detailed outline of the general procedure to
follow in using the maskcopy.
9.1 Check
controller Settings
9.1.1 Check the status
of the controls on top of the machine. It should be in the following condition.
|
Controls |
Settings |
|
Lamp Servo Mode |
|
|
N2 Purge Time |
0 |
|
Vacuum Pump Down Time |
2.0 on
Black Knob |
|
Vacuum Pump Down Time |
2.5 on
Red Knob |
|
Plate Separation Time |
3.5 |
Note: Do not adjust the settings on the top panel of the copier. They have been calibrated for optimum results. Check these settings against those posted on the front panel of the copier.
9.2 Copy 2.5 and 5 Masks
9.2.1 Check whether the
door fixture is the right size for the mask you want to copy. If not, see the
door exchanging procedure below (Section 9.4).
9.2.2 Push MAIN POWER
button.
9.2.3 Set exposure time
and intensity knobs as the above chart (Section 8.1.1 or 8.1.2).
9.2.4 Open the vacuum
chamber door.
9.2.5 Use the nitrogen
gun to blow any dust off the master plate.
9.2.6 Load the master
plate with the emulsion or chrome side facing out on the master anvil.
9.2.7 Push MASTER
VACUUM button. Vacuum gauge should read 25 or higher.
9.2.8 Push FAST PUMP
button. For 5 mask, push the MASTER HOLD button also.
9.2.9 Push EMULSION
button in for exposing emulsion plates or the EMULSION button out for exposing
chrome plates.
9.2.10 Load the copy plate against the three
metal pads with the emulsion or photoresist side up.
9.2.11 Use the nitrogen gun to blow off any dust
particles on the master and copy plates.
9.2.12 Close the vacuum chamber door. The
automatic cycle light will come on and the automatic cycle will start the
vacuum. The COPY vacuum gauge, MASTER vacuum gauge and the MAIN CHAMBER gauge
should read 25 or higher. When the vacuum is good, press the AUTOSTART button
twice to start the exposure.
Option: The
9.2.13 When the exposure is finish, the output
light on the OMRON gauge will go off. Press LATCH RELEASE button and the CYCLE
ON light will go off. Wait a minute, and then open the chamber door slowly. If
the copy plate sticks to the master plate, use a N2 gun and generally blow
between the two plates. Remove the copy plate.
9.2.14 Turn off the MASTER VACUUM, FAST PUMP and
MASTER HOLD buttons. Remove the master plate and close the chamber door.
9.2.15 Turn off the MAIN POWER.
9.3 Changing Door
Fixtures
9.3.1
Open the door.
9.3.2
Loosen the two knobs on the side of the chamber door.
9.3.3
Move the 2 black brackets out of the slots.
9.3.4
Close the door.
9.3.5
Grab the top and bottom of the door with both hands and lift it
out.
9.3.6
Slide the new door in, making sure it rests flat against the
copier.
9.3.7
Open the door.
9.3.8
Insert the 2 black brackets back into the slots of the chamber
door.
9.3.9
Tighten the knobs.
10.0 Troubleshooting Guidelines
If you suspect that the copier is not
operating properly, push the MAIN POWER button out to turn off the power on the
whole system.
If you need to change plates or exposure
in the middle of a cycle, the LATCH RELEASE button may be used. It will stop
the cycle and release the latch holding the door.
11.0 Figures & Schematics
N/A
12.0 Appendix
Mask Polarities
12.1 Emulsion mask with a clear field (black boxes) when copied
over to another emulsion plate gives a dark field emulsion mask (with windows).
12.2 A clear field emulsion mask copied over to chrome will give
a clear field chrome mask; when (the chrome copy) copied back over emulsion, it
will give a dark field emulsion copy.
12.3 A clear field emulsion mask copied over to chrome will give
a clear field chrome copy. (Copied over to chrome once more will give another
clear field copy.)
12.4 A clear field emulsion mask copied over to chrome will give
a clear field chrome copy.
12.5 A
dark field chrome mask copied to emulsion gives a clear field emulsion mask
with boxes. The emulsion copy copied back over to chrome will then give a clear
field chrome mask.
12.6 A
clear field chrome mask copied over to emulsion will give a dark field emulsion
mask.
12.7 A
dark field chrome mask copied over to chrome gives a mirror image dark field
chrome mask. The mirror image dark field chrome mask copied back over to
emulsion will give a clear field emulsion mask with the same orientation as the
original.
12.8 A
dark field mirror image emulsion mask copy copied over to emulsion again will
give a clear field emulsion. The Clear field mask copy copied to another
emulsion plate gives a dark field mirror image.
12.9 A dark field chrome mask copied over to iron oxide will give a dark
field iron oxide copy.
Note: Whenever a mask is copied, the copy (mask)
will always be a mirror image of the original mask pattern.