Chapter 6.01
Hummer Sputtering System
(hummer)
Located in
145 Cory Hall
(For SEM
Samples Only)
1.0
Title
Hummer Sputtering System
2.0
Purpose
It is to reduce the charging up of the samples by the e-beam of the scanning electron microscope.
3.0
Scope
The Hummer Sputtering System is
for depositing a thin layer of gold- palladium on the SEM samples.
4.0
Applicable Documents
4.1 Operation procedures posted on the wall by the Hummer Sputtering
System.
4.2 Hummer Sputtering System Manual is available in the office.
5.0
Safety
NEVER LEAVE THE CHAMBER UNDER VACUUM WHEN THE SYSTEM IS NOT
IN USE!
6.0
Statistical/Process data
6.1 Problem and comment section under equipment section of the wand.
6.2 Enable message for hummer.
7.0
General System Overview
The Hummer Sputtering System uses the Ar gas to
generate the plasma for sputtering gold- palladium.
The system should be in the following status before operate:
|
Main Power Switch High Voltage Switch Voltage Control Process Control Vent Valve Mode Selector Main Argon Cylinder Valve Coat/Pump Switch Argon Gas Line |
OFF OFF 0 MANUAL CLOSED D.C. PLATE On (always) PUMP ON |
8.0
Equipment Operation
8.1 Enable the (hummer) system using the wand.
8.2 The roughing pump should be on.
8.3 Open the chamber by lifting the hinged top plate. Place your specimens on the pedestal and
replaced the top plate.
8.4 Ensure that the Coat/Pump switch is set to Pump.
8.5 Plating Procedure:
8.5.1
Place mode switch to D.C. Plate.
8.5.2
Argon valves should be turned on the gas line. (DO NOT adjust the regulators. DO NOT change the pressure.)
8.5.3
Turn on the Hummer main power switch. Pump until the pressure is between 20 mtorr
to 50 mtorr. (If it doesn’t happen,
switch the valve to “Coat” for a while.
This will allow argon gas to flush any water molecules or nitrogen out
of the chamber. The pressure should be
approximately 200 mTorr. You may open
it up to about 500 mTorr for flushing the chamber. DO NOT use this knob as a shut off valve!)
8.5.4
When the chamber pressure reaches 20 mtorr to 50 mTorr, turn
the valve from “Pump” to “Coat” and allow the pressure to stabilize at 200
mTorr.
8.5.5
Turn high voltage power on.
8.5.6
Turn high voltage dial until the current is at 25 mA. (You should see a purple plasma.)
8.5.7
Turn the timer dial to the desired time setting and switch
to “Auto”. A thickness of 50A to 100A
is sufficient for most SEM viewing purposes.
Generally, 1-2 minutes coating is enough. Pulse mode can reduce the temperature of the deposition on the
samples. On the pulse mode, the timer
only counts the time when Cycle is on.
8.5.8
After the coating time has elapsed, the plasma will shut off
automatically.
8.6
Turn off and unload procedure:
8.6.1
Turn high voltage dial to 0.
8.6.2
Turn off high voltage switch.
8.6.3
Switch the valve from “Coat” to “Pump”.
8.6.4
Set the timer switch back to manual.
8.6.5
Turn off the Hummer system main power switch.
8.6.6
Vent the chamber; remove your samples.
8.7
Close the vent valve and the top lid.
8.8
Disable the system.
(The vacuum pump should be off.)
9.0
Troubleshooting Guidelines
DO NOT leave the Hummer operating
unattended! If you believe something is
not working right, write up a Fault report- DO NOT try to “fix” it.
Note: Never leave the chamber under vacuum. It will suck all the pump oil out of the pump in the event of a power
failure. Leave the chamber vented!
Rev. 00 – 01/02, K. Chan
Rev. 01 – 12/06, K. Chan – Equipment moved to Room 145.