Chapter 7.14
Ionmill Procedures
(ionmill)
1.0
Title
Ionmill Procedures
2.0
Purpose
3.0
Scope
The
following settings work well for the ionmill. You may need to modify them for
your particular task (e.g., to achieve uniformity on large samples, see note
under Step 9.5.3). Nevertheless finding a set of suitable conditions and
staying consistent will enable you to predict etch rates and required mill
times for a variety of materials. For (0001) sapphire (c-axis), the following
settings will give a depth of ≈ 0.65 µm in 45 min. Shipley 1818
photoresist (soft baked for 20 min at 90°C and hard baked for 20 min at 120°C)
will mill at 2.0 µm/h, so samples coated with the normal procedure (30 s at
5500 rpm which yields ≈ 1.8 µm thick photoresist) should not be ion
milled any longer than 45 min since overetching may begin to round corners (of
course over-etching may be desired in special circumstances). Glass will etch
at ≈ 1.5 µm/h using the following settings.
4.0
Applicable Documents
5.0
Definitions and Process Terminology
6.0
Safety
7.0
Statistical/Process Data
8.0
Available Process, Gases, Process Notes
9.0
Equipment Operation
9.1
General
Preparation
9.1.1
Check cryo pump
temperature (it must be < 18 K).
9.1.2
Check if the vacuum
pressure is low. It should read 0 mtorr on the digital gauge. Turn on the ion
gauge filament via the switch on the ion gauge controller. Base pressure for
the ionmill should be in the low 10-6 torr range on the ion gauge.
9.1.3
Check the argon
pressue: There is a small, round gauge mounted on the Ar feedline, near the
hoist. Ar pressure should be 3-5 psi.
9.1.4
Ion milling transfers
considerable heat to the substrate being etched. When polymer masks are used,
some method of heat transfer to the
rotating water-cooled substrate table is required.
Cool
Grease™ is a thermally conductive paste suitable for ion milling. A jar of Cool
Grease™ is available at the ionmill. If
it needs to be replaced, lab members may check out additional Cool
Grease™ from the office. Information about Cool Grease™ can be found at: http://www.aitechnology.com/
Loading
Sample into Ionmill
9.1.5
Close high vacuum valve and wait for pop (clank) to insure
it is closed. It may take ≈ 30 s for the pop to sound, and its loudness
has changed over time, but it is very important to wait for this.
9.1.6
Vent chamber and wait for digital gauge to stabilize at
atmospheric pressure (≈ 700 - 760 torr) to insure complete ventilation.
9.1.7
Raise hoist and turn vent valve off. Make sure the gasket
(giant O-ring) stays in place, if not put in place. It is usually a good idea
to always push it down around the entire chamber wall even if it does not come
undone completely.
9.1.8
Apply Al paste to stage so that most of plate (or larger
wafer) will have contact with the paste.
9.1.9
Push on plate (or large wafer) so that paste is smashed
around. Blow off surface with N2 gun.
9.1.10
Lower the hoist. Make sure the stage apparatus does not hit
the chamber walls during lowering.
9.2
Pumping Down
Chamber to Sufficient Vacuum
9.2.1
Open roughing valve and
pump down to 100 and 200 mtorr on pressure gauge.
9.2.2
When pressure gets
between 100 and 200 mtorr close roughing valve,
make sure pressure is stable (i.e., there are no major leaks) then open high
vacuum valve. (In the old manual it states to rough the chamber to 200 - 400
mtorr; however, I routinely rough vacuum furnaces to less than 100 mtorr before
starting a turbo pump or diffusion pump, and so I have selected 100 - 200 mtorr
as a compromise.)
9.2.3
Turn ion gauge on and
allow system to pump down to < 4 × 10-6 torr.
Depending on your application,
you may need to let the system pump down to a better vacuum; using the
procedure in the next section and Steps 9.7.6 and 9.7.7 to flush the chamber
with argon can help attain a better vacuum. However, recently I’m not sure the
ionmill will pump much better than about 2 × 10-6 - torr, even after leaving it overnight.
9.3.4 While
waiting for the system to pump down, turn on table rotation and coolant. Verify
that vacuum is not lost due to rotation, or a water leak. Also, verify water
flow. Previously, the rotation and water could be turned on at any time;
however, after some modifications to the equipment, it appears as if they will
not come on until the high vacuum valve is opened.
9.3
Preparing
Chamber with Proper Argon Pressure
9.3.1
Make sure all (not the
limiters though, these should be all the way up) knobs for ion beam controls
are completely turned down, and that all devices are turned off. Some people
are very sloppy about shutting the machine down.
9.3.2
Turn main power on
(breaker).
9.3.3
Make sure argon pressure
is set to 2 × 10-5 torr.
9.3.4
Turn argon power and
filament on. (Make sure the pressure button for the Ar-ion gauge is depressed; i.e.,
not the deviation, calibration, or some other setting, unless of course you
prefer to monitor these.)
9.3.5
Make sure valve closed
light is on.
9.3.6
Switch from valve to
control mode by pressing the control/close button.
9.3.7
Wait for argon pressure
to get near 2 × 10-5 - torr, then turn
pressure setting up to 8 × 10-5 torr.
It will usually take about a minute for the automatic valve to engage and for
the pressure to reach the initial set point of 2 × 10-5
- torr.
9.3.8
Wait for pressure to
equilibrate at 2 × 10-5
torr; meanwhile, make sure the shutter is closed.
9.4
Setting Ion Beam
Controls
9.4.1
Turn the isolation transformer on (lights will come on if it
is indeed on). If it does not turn on, try flipping the lever slowly.
9.4.2
Turn cathode knob up to a current of 19 A, which should
correspond to a power of 150 W and 7.5 V.
9.4.3
Turn magnet on and up to 0.4 A.
9.4.4
Turn the arc up to 45 V; this should correlate with about 1
A. You can now adjust the cathode slightly to get 1 A reading on the arc.
9.4.5
Make sure all three (cathode current, magnet current, &
arc current) are stable.
9.4.6
Turn suppressor on, and up to 250 V.
9.4.7
Turn accelerator main on, then press HV on button. (You may
need to turn dial up and down in order to get the HV to turn on properly.)
9.4.8
Adjust the acceleration potential to 1000 V. This should
correspond to about 1 0 A/cm2 on the
current density display. (At this time, if the suppressor voltage oscillates,
it is due to the degassing of species inside the chamber. By waiting a short
while these oscillations should stop, but they are not a major concern.)
9.4.9
Adjust shutter until maximum is read on current density
display.
9.4.10
Turn neutralizer on, adjust until there is zero readout for
current density and target current. This will usually be in the vicinity of
10-15 V.
9.4.11
After these adjustments, the acceleration potential should
be about 900-1000 V. The arc current will be ≈ 1.1 A.
9.5
Milling Samples by
Opening Shutter for Desired Time
During milling, make
sure all gauges are stable. In particular watch the target current. Small
little spikes in the current are okay; however if a current > 0 develops,
compensate with neutralizer. Also, keep an eye on the suppressor, sometimes it
extends past 250 V (i.e., off the gauge range), so you will need to adjust it.
Fluctuations in the controls tend to occur in the early stages of milling. You
should keep an eye on the controls for the first 20-30 min of milling, but if
you need to mill for an extended time (say a few hours) check it every 20-30
min after the initial time. Close shutter after desired time period.
9.6
System Shutdown
9.6.1
Turn down the acceleration all the way, push HV off button,
and then turn off main.
9.6.2
Turn down the arc all the way.
9.6.3
Turn cathode down all the way.
9.6.4
Turn the magnet, suppressor, and neutralizer off in any
order. Always turn one knob down, then turn that control off before proceeding
to the next control.
9.6.5
Turn off the isolation transformer.
9.6.6
Turn the argon pressure setting to 2 × 10-5 torr.
9.6.7
Switch to valve closed mode (press control/closed).
9.6.8
Turn off the filament and then the power to the argon
controller.
9.6.9
Let system cool for 5-10 min, meanwhile, turn off table
rotation.
9.6.10
Turn off master power (breaker), coolant, and filament.
9.7
Removing Sample
and Leaving Chamber under High Vacuum Conditions
9.7.1
Close high vacuum valve (again wait for pop to insure it is
closed).
9.7.2
Vent chamber.
9.7.3
Raise hoist and turn off vent, make sure gasket (giant
O-ring) is in place after hoist has risen.
9.7.4
Remove sample/plate.
9.7.5
Clean stage with razor blade and wipe with acetone or
isopropanol. Also, clean off the copper plate if you used it.
9.7.6
Lower hoist.
9.7.7
Open roughing valve.
9.7.8
When pressure is between 100 and 200 mtorr, close roughing valve,
then open high vacuum valve.
10.0
Troubleshooting Guidelines
10.1 Problem: Neutralizer not working.
Solution: Replace broken neutralizer wire. The
.008" Tungsten wire is located in the sea locker drawer lableled ionmill
in the back of the old lab. Cut wire to length adequate to wrap around posts.
First, unloosen nuts and remove old wire. Wrap with new wire around post twice
and stretch to next post, and wrap again. Secure nuts to wire. Cut off excess
wire.
11.0
Figures and Schematics
12.0
Appendices
Rev. 00 – 9/02, X. Meng
Rev. 01 – 10/05, R. Hamilton – Changes on 9.1 General Preparation.
Rev. 02 – 2/06, R.
Hamilton – Sections 9.1.4 and 9.1.5 deleted and replaced with new information
on Cool Grease™.
Rev. 03 – 2/07, J. Donnelly
– Added one problem in Section 10.0 – Troubleshooting.