Chapter
8.23
1.0
Title
UVSCOPE
Operation
2.0
Purpose
This manual describes the general
operation of UVSCOPE and its interface with XCAPtm Interactive Image
Analysis Software.
3.0
Scope
This manual includes the procedures for Bright Field, Dark Field,
Interference contrast, and UV sub-micron operations. It also includes the procedures of grasping the microscopic
images using XCAPtm Interactive Image Analysis Software and
transferring the images to the user’s computer workstation.
For lamp replacement, refer to Section 4.2.
For details of XCAPtm Interactive Image Analysis V2.0/2.1, refer to
Section 4.3.
4.0 Applicable
Documents
4.1
Revision
History
4.2
Leica
INM100 Universal Inspection Microscope Operation Manual (Available at the
UVSCOPE and in the office)
4.3 XCAPtm Interactive Image Analysis V2.0/2.1 Reference Manual (Available online and in the office)
5.1 Ergotube - A combination of binocular tube and video tube. It can be adjusted for loss less beam splitting into 100% eyepieces (position I), 50% eyepieces-50% video (position II), or 100% video (position III).
5.2
Bright Field Image – Microscopic image of reflected light.
5.3
Dark Field Image – Microscopic image of scattered light.
5.4
Differential Interference Contrast Image – Special
application for inspection of surface topographies.
5.5
UV Image – Microscopic image of reflected Ultra Violet Light
(365nm) for inspection of sub-micron features.
6.0 Safety
6.1
All general Microlab safety guidelines apply when operate
UVSCOPE.
6.2
HOT SURFACE HAZARD: Lamp house surface is hot and may cause
burn.
6.3 EXPLOSION HAZARD: Hg and Xe Lamp may explode and can cause severe injury if not secured in explosion shielded housings
7.0 Statistical/Process Data
N/A
8.0 Available Processes, Gases, Process
Notes
N/A
9.0 Equipment Operation
9.1
General UVSCOPE Operation – Bright Field Operation
9.1.1
Enable
UVSCOPE on the Wand. Turn on the microscope and the bright light power by press
the main switch (on the lower right side of the microscope) to “on”
position. If you are going to use the
UV light, turn on the UV light power supply unit (ebq 100, Netz Power). It needs 20 minutes to warm up.
9.1.2 Lower the
stage with the coarse focus knob. Move
the stage to the right and toward you by moving with the two knurled knobs
under the stage on the right hand side. Put your wafer on the center of the
stage. The stage can be rotated 360 degree if needed. Move the stage back to the position where you can examine your
wafer. Do not try to put your wafer
directly in between the objective lens and the stage. You may scratch or damage
the objective lens and your wafer.
9.1.3 Check
microscope set up.
9.1.3.1 Check the
position of the Ergotube (on the center right side). If it is pull out at position “III”, push it in to the position
“II” or “I”.
9.1.3.2 Check the
Contrast turret (on the right center side of the microscope) is set at BF. If not, turn it until BF is locked at the
position.
9.1.3.3 Check the
Interference prism turret (on the front side of the microscope) is set at “H”. If
not, turn it until “H” is locked at the position.
9.1.3.4 Check the
UV light shutter (between the microscope and the UV light housing) is at the
bright light position.
9.1.3.5 Check the
Bright light shutter handle (on the left back side of the microscope) is to the
“in” position.
9.1.3.6 Check the
incident light aperture. The “man. AD” button (on the lower left side) should
be red. If it is white, push it once to
change it to red.
9.1.4 Select the
suitable objective lenses using the “+” and “–“ buttons located on the lower
left hand side of the microscope. Push the “+” button for higher magnification
lens and “-“ for lower. Do not turn the
objective turret manually. It is easier to focus starting with lower power
objective lens.
There are six lenses on the UVSCOPE:
HC PL FLUOTAR 5X Red ring
HC PL FLUOTAR 10X Yellow
ring
HC PL FLUOTAR 20X Green ring
HC PL FLUOTAR 50X Blue ring
PL APO 150X White
ring
UV 365nm 150X White ring (smaller size)
9.1.5 Adjust the
lamp brightness using the thumb wheel located in the lower front of the
microscope.
9.1.6 Choose the
lowest magnification lens. Raise the
stage while focusing. Focus on your
wafer using the coarse focus drive and turn slowly while looking through the
eyepieces. With the coarse focus drive (outer knob), the stage can be fast
positioned in height. The fine focus (inner knob) has two different gears. When
it is pushed to the left side, one rotation equals 0.1mm. When it is pushed to the right, one rotation
equals 0.02mm. This is extremely useful for high power objective lens.
9.1.7 Perform
stage stop position set up.
9.1.7.1 This is
needed to prevent damage to the sample due to the stage position being too
high; the upper stage stop position can be set with the limiting wheel (black
ring around the coarse focus knob).
9.1.7.2 Disengage
the upper limit stop by rotating the limiting wheel counterclockwise a quarter
turn. If you do not want to set the upper limit stop, proceed to 9.1.8.
9.1.7.3 Set
correct focus using 50X objective.
9.1.7.4 Engage the
upper limiting stop by rotating the limiting wheel clockwise until it is fixed.
9.1.7.5 This
procedure must be repeated whenever the sample thickness is changed.
9.1.8
Adjust eyepiece distance and correct focus for un-even
eyesight if desired.
9.2
Dark Field Operation
9.2.1 Follow
Procedure 9.1 to obtain the Bright field image first.
9.2.2 Turn the Contrast
turret (located on the center right side of the microscope) to “DF”
position. Adjust the lamp brightness
and focus if needed.
9.2.3
To return to the Bright Field image, turn the Contrast
turret to “BF” position.
9.3
Interference Contrast Operation
9.3.1 Follow Procedure
9.1 to obtain the Bright field image first.
9.3.2 Turn the
Contrast turret (on the center right side of the microscope) to “DIC” position.
9.3.3 Turn the
Interference prism turret (on the front side) to the appropriate position:
Position 1 HC PL FLUOTAR objectives (5 to 50X)
Position 2 same as position 1
Position 3 PL
APO objectives (150X)
9.3.4 Adjust the
small knurled knob (on the right side of the Interference turret) to obtain
optimal interference contrast.
9.3.5
To return to the Bright field image, turn the Interference
prism turret to “H” position first. Then turn the Contrast turret to “BF”.
9.4
Viewing the image using the SONY monitor
9.4.1 Check the
position of the Ergotube (on the center right side). If it is pushed in at the position “I”, pull it out to the
position “II” or “III”.
9.4.2 Set the
A/B switch box to “A” for Bright Field, Dark Field, and Interference Contrast
Images. (Switch to “B” for UV image.)
9.4.3 Turn on
the SONY monitor. Select Channel (line)
A. The settings (brightness, contrast, and etc.) are preset. To turn off the
pre-settings for manual adjusting, press “manual” button and follow the
instruction on the screen.
9.4.4
You may need to adjust the microscope lamp brightness and/or
focus to obtain satisfactory images.
9.5
UV image Operation (can only be displayed on the SONY
and PC monitors)
9.5.1 Follow
Procedure 9.1 to obtain the Bright Field image of the area of interest. Switch the objective to UV 365nm (150X).
9.5.2 Change the
UV light shutter position (between the microscope and the UV light housing) to
UV position.
9.5.3 Pull out the
Bright light shutter handle (on the left back side of the microscope).
9.5.4 Check the
position of the Ergotube (on the center right side). If it is pushed in at the position “I”, pull it out to the
position “II” or “III”.
9.5.5 Set the
A/B switch box to “B” for UV image.
9.5.6 Turn on
the SONY monitor. Select Channel A. The
settings (brightness, contrast, and etc.) are preset. To turn off the
pre-settings for manual adjusting, press “manual” button and follow the
instruction on the screen.
9.5.7 You may
need to adjust the UV Video Camera Shutter setting (black push button device
sits on top of the microscope) and/or microscope focus to obtain satisfactory
images.
9.5.8
To return to the Bright Field image, follow Procedure 9.1.3
and set the A/B switch box to “A”.
9.6
XCAP and FTP Graphic File Transfer
9.6.1 Check the
position of the Ergotube (on the center right side of the microscope). If it is pushed in at the position “I”, pull
it out to the position “II” or “III”.
9.6.2 Turn on
the UVSCOPE PC Workstation next to the microscope. It is running Window NTTM OS. The login is “leica’ and
the password is “LEICA” too (case sensitive).
9.6.3 Double-Click
the “XCAP for Windows” icon to start the program.
9.6.4 Select the
drop down menu “PIXCI®”; select “PIXCI® open/close”. In the pop-out window,
click “open” button. A window with previous grasped image will pup-out. If no
image (dark screen), then refer to troubleshooting guidelines Section 10.
9.6.5 Select the
proper camera connection in the setup window.
Top connection is for Bright Light images and bottom for the UV. Click
“snap” or the “camera” icon to grasp new images.
9.6.6 You can
save the grasped images in several graphic formats on a floppy diskette or the
hard drive temporarily. If you save the images on the hard drive, please use
SSH Secure Shell to transfer the files to your own workstation account. The
users’ files will be purged to protect the hard drive.
9.6.7
To use the SSH Secure Shell program, click the “start”
button on the bottom menu bar. Click
the “programs”, choose the “SSH Secure Shell”, then the “Seccure File Transfer
Client” option. Click the “Quick
Connect” icon. Select your host computer and account. Enter your password for
the account. Drag the graphic files to your host computer directory.
9.7
Fluorescence Operation (FL) – Not Available
9.8
Shutdown Procedures when you are done.
9.8.1 Return the
microscope to Bright Field state by following Procedure 9.1.3.
9.8.2 Turn down
the lamp brightness to minimum.
9.8.3 Turn off
the microscope main switch and the UV light power supply (ebq100).
9.8.4 Turn off
the SONY monitor and the PC workstation monitor. Leave the PC workstation on.
10.0
The wafer stage does not move up for focusing.
Check the stage upper limiting
stop is not engaged (Procedure 9.1.7).
10.1 No Bright
Field, Dark Field, or Interference Contrast Image
10.1.1 Check
there is light coming out of bright light lamp housing (left rear side of the
microscope). If there is no light coming out, the lamp needs to be changed.
Report fault on the Wand.
10.1.2 Double
check the microscope setup by following Procedure 9.1.3.
10.2 No UV
Image
10.2.1 Check
there is light coming out of UV lamp housing (rear side of the microscope). If
there is no light coming out, the lamp needs to be changed. Report fault on the
Wand.
10.2.2 Double
check the UV image operation Procedure 9.1.3.
10.3 No image
on the “SONY” monitor and/or XCAPTM
10.3.1 Check
there is power to the monitor.
10.3.2 Double
check Procedure 9.4 for the “SONY” monitor, Procedure 9.6 for XCAPTM.
10.3.3
The light from the microscope may be too low for the video camera.
Manually adjust the incident light aperture (see instruction in the appendix).
11.0 Figures & Schematics
Refer to Leica INM100 Universal Inspection Microscope Operation Manual
12.0 Appendix
The incident light aperture can
affect the resolution and contrast of the microscopic image. The UVSCOPE is
equipped with a variable iris aperture diaphragm which is factory aligned for
the best performance. In some cases, the application requires an overwriting of
the factory setting.
To adjust the aperture diaphragm
manually, push the red [Man. AD] button (on the left side of the microscope)
once. It will turn white/pink. Then adjust the aperture diaphragm using the
[AD] rotary wheel under the [Man. AD] button. Pushing the [Man AD] button
again, it will turn red and the diaphragm return to the automatic mode.
In the automatic mode, the
diaphragm will be adjusted to the preset positions. When changing from Bright
Field to Dark Field, the diaphragm will open automatically. Changing from Dark Field to Bright Field,
the diaphragm will return to its previous position.