Chapter 8.41

JEOL 6400 SEM and

Nanometer Pattern Generation System

(jeol107)

 

1.0          Title

JEOL107 – JEOL series 6400 Scanning Electron Microscope (SEM), and Nanometer Pattern Generation System (NPGS)

2.0          Purpose

JEOL107 is a scanning electron microscope that can be used for imaging of samples up to 4” in size, with up to 300,000X magnification. It can also be used for electron beam lithography, using the Nanometer Pattern Generation System (NPGS).

3.0          Scope

This manual will describe the basic operation of the JEOL 6400 microscope, both for imaging and for its use with the NPGS E-beam lithography system.

4.0          Applicable Documents

4.1                JEOL 6400 manual (located in Cory 107, on the shelf above the microscope).

4.2                Nabity NPGS manual (located in Cory 107, on the shelf above the microscope).

4.3               Further information on NPGS is available on the JC Nabity website at: http://www.jcnabity.com.

5.0          Definitions & Process Terminology

5.1                SEM: Scanning Electron Microscopes are used for imaging with very fine (nm) resolution. This equipment is also configured for Electron Beam Lithography (EBL) using the NPGS software and hardware.

5.2               NPGS: The Nanometer Pattern Generating System allows the user to pattern a sample from a computer layout design (Design Cad). The NPGS controls the beam position and blanking during the writing phase.

6.0          Safety

The SEM operates under high vacuum. Care must be taken during loading/removing of the sample, and during adjustments being made on the column, to avoid loss of vacuum to the column and damage to the filament. Samples must be properly prepared for minimal outgassing and chamber contamination, and handled with gloved hands only.

7.0          Statistical/Process Data

Enable messages for JEOL107.

8.0          Available Process, Gasses, Process Notes

8.1                Standard Imaging

The JEOL 6400 scanning electron microscope (SEM) is a high-resolution SEM. It can provide beam voltages ranging from 0.2kV to 40 kV and beam currents from 10 picoamps to 10 microamps. It offers high performance and low noise at low accelerating voltages, making it useful for IC and photoresist evaluation. Resolution of 3.5 nm is attainable, and available magnifications range from 10 X to 300,000 X. The cathode is a high-brightness lanthanum hexaboride (LaB6) source. The SEM is equipped with two-inch and four-inch airlocks and a Faraday cup for beam current measurements. The sample stage is computer-driven.

8.2    Electron Beam Lithography

The JEOL 6400 is equipped for electron-beam lithography (EBL). A beam blanker with a rise time of 3 microseconds is installed. The beam blanker and scan coils are linked to a computer with a commercial pattern generation system, the Nanometer Pattern Generation System (NPGS) from J. C. Nabity Lithography Systems. This system is designed to generate patterns ranging in scale from nanometers to millimeters. Patterns are designed using commercial CAD software (DesignCAD) and then assigned exposure parameters using the Make Run File (mrf) command. The resulting run files are then written by calling the NPGS program. NPGS makes calls to Pattern Generation (PG) and Alignment (AL) programs as necessary. Both PG and AL can be run separately if desired. PG writes the pattern by controlling the X-Y scan coils and beam- blanking of the SEM. Resolution is 16-bit for the beam position and 12-bit for the beam blanking. By using AL, patterns may be aligned to existing alignment marks without exposing the pattern area. The user confines the beam raster to designated windows. The image from the windows is displayed both on the CRT of the SEM and at the monitor of the pattern generation computer. Overlay marks are simultaneously displayed on the computer screen, and can be aligned with marks on the sample. Rotational and scaling information from the alignment process is used to generate a transformation matrix. PG can use this matrix subsequently when generating patterns.

9.0          Operating Procedure

9.1        Daily Maintenance/Pre-Use Checks

9.1.1        When not in use, the JEOL 6400 is normally left with the high voltage turned on but set to 0 kV. This leaves the LaB6 filament pre-heated with ~ 1.0 A running through it. Keeping the filament warm maintains its cleanliness, and allows for faster start-up when using the SEM. Note that all the controls required for standard sample observation and EBL are described below. Under no circumstances should users operate any other controls without first consulting staff. This includes the entirety of the vacuum system, with the exception of the sample airlocks.

9.1.2        Prior to using the SEM you should check that all is in order. First, check that the gun pressure is acceptable. The gun is pumped by an ion pump whose power supply is located to the right of the SEM control cabinet. The power lamp should be illuminated. The meter range switch should be on the 500-microamp scale (change scales if necessary) and the meter should read toward the low end of the scale, typically about 70 microamps. This meter reads the ion pump current, which is proportional to the gun pressure. If the meter reads above 250 microamps, the pressure is too high for the SEM to be used. Report the problem on FAULTS.

9.1.3        Verify that the filament knob is turned fully counterclockwise and that the filament meter light is on. The filament button FIL toggles the meter between filament and emission current measurement. If FIL is not lit, you    are looking at the emission current, and should toggle to FIL; the meter should then read ~1.0 A.

9.1.4        Verify that Vib is closed: Vib LED is off, Vib button is in.

9.1.5        If the meter light is not on, the gun pressure is not acceptable. If both FIL and the meter light are lit but there is no filament current, the SEM is not usable. Report the problem on FAULTS.

9.1.6       Check that the accelerating voltage is set to 0 kV and the secondary electron detector is off using the EOS menu.

9.2               Keyboard and On-screen Menus

Most of the functions of the JEOL 6400 can be accessed through keyboard and menu operations. This includes most of the knob and button controls, with a few exceptions (e.g., the filament current knob). In addition, much of the fine-tuning required for successful lithography can only be accomplished using keyboard control. The various menus through which control takes place are displayed only on the right-hand (RH) CRT. The left-hand (LH) CRT is used primarily for viewing. A cursor is displayed on one of the two CRTs at a time; the CRT in which the cursor appears is active. The cursor can be toggled back and forth between the two CRTs by pressing the escape (ESC) key.

9.2.1        Computer Keyboard

The most useful keys are escape (ESC) mentioned above, BREAK, the return key, and some of the pre-defined function (PF) keys. Also useful are insert (INS), and the plus (+) and minus (-) keys. Uses of these keys will be described below.

9.2.2        Basic Screen

The basic screen is always displayed on the LH CRT; it is   also displayed on the RH CRT when the menus are hidden. This screen consists of two information lines customarily located at the bottom of the screen. When a given CRT is active (i.e. the cursor is in it) and is displaying the basic screen, its information lines can be alternately   shown or hidden by pressing the BREAK key. The information displayed in the lines includes a micron bar and size, a title, the accelerating voltage, the magnification and the working distance. The pre-defined function key PF1 brings up the basic screen on the RH CRT regardless of which screen is active.

9.2.3        Electron Operating System (EOS) Menu

PF2 brings up the EOS menu on the RH CRT regardless of which screen is active. The EOS menu consists of three screens: EOS-1 (optics), EOS-2 (mode) and EOS-3 (image). When the RH CRT is active and EOS is displayed, pressing the 1, 2, or 3 keys will call up the corresponding screen. Most of the SEM functions required for viewing or lithography can be controlled though the EOS menu. To change a parameter, make the RH CRT active using the ESC key. If the item you want to change displays a set of possible choices (e.g. the raster scan speed), position the cursor over your choice using the arrow keys. Then press the return key to select the item. If instead the item is numeric (e.g. the accelerating voltage), position the cursor over the number and press the INS key. A white command window will appear on the RH CRT. Enter the new setting and press return to make a change. You can also increment or decrement a numeric setting by positioning the cursor over it and pressing the + or - keys.

9.2.4        EOS Screens

Settings available in the EOS screens are indicated in the list found in appendix 1 (Section 12.1). Ranges are shown for numeric items when appropriate. When control over an item is not needed for basic use, only the standard setting of that function is listed. Settings not needed for basic use are not listed, and should not be used without consulting staff. Standard settings of the other items are indicated by asterisks.

9.2.5        MEMO Menu

The MEMO menu is called up by pressing the PF5 key. This menu lists ten sets of pre-defined operating conditions; the chosen settings for various SEM parameters are listed in the lower half of the screen. To load the selected set of operating conditions, select the title of the set in the upper half of the screen using the arrow keys. The press   "l" for load. The listed set of conditions will immediately be loaded into the SEM.

Note: Loading a set of operating conditions will result in immediate application of an accelerating voltage! DO NOT load a set unless you are ready for the listed conditions to come into effect immediately. See Section 9.3: System Operation below.

Do not change the contents of any of the pre-defined states; they are provided for the convenience of all users.

9.3        System Operation

9.3.1        Sample Preparation

Samples can be most sizes or shapes up to a whole 4" wafer. Samples must be clean and dry, i.e., free from moisture, oils, grease, and particles that could contaminate the vacuum system. If the chamber becomes contaminated, the contaminants can be "ironed on" to your sample by the beam. Once this happens, it is usually impossible to remove the contaminants from your sample.

9.3.1.1   Gloves must be worn when handling samples and sample holders. Baked photoresist can be inspected in the jeol107. Note that nonconductive materials are subject to "charging" and for best results should be coated with a thin layer of gold. For cross-sections, cleave the sample through the area of interest.

9.3.1.2   The sample should be as small as possible, but large enough that it can be securely held by the sample holder. If the sample is to be coated with gold, it should be coated after cleaving. For cross sections, use the slotted sample holder; these holders have a set screw for securing the sample. Be careful not to over tighten the screw, or you may break your sample. The "Z" axis knob on the stage is calculated from the sample holder top. If the sample must be secured to the holder, use SEM-compatible (vacuum compatible and conductive) carbon paint.

9.3.1.3   After attaching or securing your sample to the sample holder, the sample holder is mounted into the larger dovetail holder. This dovetail holder will slide or "dovetail" over the stage in the SEM chamber. For cleaved cross-sections, it is recommended that you mount the sample so that the cleaved edge is parallel to the longer side of the dovetail holder.

9.3.2        Loading the Sample and Pumpdown (2 Inch Airlock)

The sample is loaded using a threaded rod; this rod is mounted through the center of a thick, circular viewport. The viewport allows you to visually guide the dovetail holder onto the stage and visually assess the orientation of the sample for tilt and rotation before viewing on-screen, where orientation is not as obvious.

9.3.2.1        Thread the screw at the end of the sample-loading rod into the dovetail holder, and hook the clip attached to the viewport over the screw. The clip secures the sample holder to the viewport until you are ready to move it onto the stage; if you don't use the clip, the sample and rod will move forward into the airlock door, possibly damaging your sample or the door! If the rod does not easily slide through the viewport vacuum seal, apply grease to the rod sparingly.

9.3.2.2        Verify that the working distance (Z height) is set to the specimen exchange distance of 39mm; the red light on the front of the airlock exchange panel will come on to indicate that the working distance is correctly set to 39mm. Also check that the fine Z micrometer is set to 2.0 mm; it is located between the X and Y micrometers.

9.3.2.3        Set the stage to its home position by typing "h" at the keyboard of the stage control computer; this should automatically set the X and Y at the home or center position (25 mm on the X and 35 mm on the Y travel counter). Manually adjust the rotation (angular and circular) counters to zero. Warning: Any operations on the optics table must be done gently, without disturbing the table. Bouncing the table even a little will crash the turbo-pump - it is magnetically levitated.

9.3.2.4        Check that the white toggle switch on the four-inch airlock is set to OTHER. The four-inch airlock is located to the left of the SEM and the two-inch airlock is in front. When the toggle is set to OTHER, the white light on the two-inch airlock should be illuminated, indicating it is the active airlock.

9.3.2.5        Check the viewport and airlock o-ring for dust, remove dust to make a good seal. Hold the viewport with the sample facing towards the SEM chamber against the o-ring seal at the two-inch airlock entrance. Press the illuminated white button on the airlock chamber and wait for the light to go out. The light is on a timer and should allow enough time for the chamber to evacuate. Improper evacuation of the airlock can cause the turbo to "dump" when you open the airlock door resulting in down time!

9.3.2.6        When the light goes out, open the airlock door by    rotating the airlock knob towards you 1/4 of a turn. Slide the door open by pulling the knob to the right, about 3". Be gentle. There is a clip in the airlock to hold the door open. Do not lock it with the knob as it will get debris on the door o-ring. The light in the SEM chamber should come on when you open this door. If the light is off, then the secondary electron detector is on. To turn off the detector, select OFF under DETECTOR in the EOS-3 menu. Improper evacuation of the airlock can cause the turbo to "dump" when you open the airlock door resulting in down time!

9.3.2.7        Carefully push the sample rod into the chamber and slide the dovetail holder onto the stage. You may have to slightly raise the sample holder in order to get it onto the stage. The sample holder should NEVER have to be forced onto the stage.

9.3.2.8        Unscrew the threaded rod from the dovetail sample holder. Make sure that the rod is completely unscrewed before retracting or you may pull your sample off the stage.

9.3.2.9        Gently retract the rod until the screw is secured in the clip attached to the viewport (you will hear a click when this happens).

9.3.2.10     Slide the door closed and turn the knob 1/4 turn away from you until the flat area on the knob is parallel to the floor.

9.3.2.11     Holding the rod near the viewport, press the white airlock button and wait for the light to go on. Make sure that you hold the rod during the venting of the airlock or the rod could fall to the floor!

9.3.2.12     Remove loading rod and place it in its holder.

9.3.2.13     Verify that the vacuum system is functioning properly. The filament meter light should still be on; this indicates the pressure is low enough to apply high voltage. Also, ensure that the gun pressure is still acceptable by checking the ion pump current. If the filament light is off or there has been a significant rise in column pressure, there may be a problem in the vacuum system. Report the problem on FAULTS. If the filament light has gone out or the gun pressure has risen above the acceptable limit, the SEM SHOULD NOT BE USED, AND YOU SHOULD LEAVE YOUR SAMPLE INSIDE. Contact staff to have your sample removed.

9.3.2.14     Never open the airlock door to the specimen chamber unless you have just pumped out the airlock. It will not preserve the vacuum and unless pumped immediately before opening, will cause the turbopump to crash, shutting down the SEM!

9.3.3        Turning on the Beam

Note: The area swept by the beam either on the screens or on the sample is called the raster.

9.3.3.1        Check that control of the raster is given to the SEM. The small metal switch in the Raster Control auxiliary panel to the right of the CRTs should be set to SEM (down). If it is set to NPGS (up) control of the raster is given to the lithography computer, and a single  spot in the center of the LH CRT will be lit. IT IS EASY TO BURN THE PHOSPHORS AT THAT SPOT IF THE CONTRAST AND BRIGHTNESS ARE TURNED UP WHILE IN NPGS MODE. Switching the raster control to SEM protects the CRT from burn-out.

9.3.3.2        Bring up the information bar in the basic screen (using ESC and BREAK), or display EOS-1 in the RH CRT. Increase contrast on screens so that display information is visible. Verify that the machine is in SEI mode (SEI button will be lit). If not, press the SEI button or select SEI under IMS CH1 in EOS-3.

9.3.3.3        Check that the Beam Blanking toggle switch is set to OFF. The Beam Blanker control is located in one of the auxiliary panels to the right of the CRTs. OFF leaves the beam unblanked, ON blanks the beam, and EXT turns control of the blanking over to the lithography computer. Normally EXT leaves the beam blanked except during an exposure.

9.3.3.4        Check that the probe current detector, which collects the beam current for measurement, is not in the beam path. The blue PCD button in the auxiliary panels should be out (unlit).

9.3.3.5        On the JEOL keyboard, set SCAN options to FAST and single line scan (LSP) by pressing the FAST and MODE buttons (buttons will illuminate). This may also be accomplished using EOS-2. The MODE button toggles between line, cursor and point options. Continuing to press MODE will eventually bring up the line scan.

9.3.3.6        Use the BRIGHTNESS and CONTRAST knobs (SEI IMAGE) to adjust the signal so that the line is positioned just above the bottom of the screen.

9.3.3.7        Apply an accelerating voltage. You can set the accelerating voltage by turning the ACCEL VOLTAGE knob, by changing ACC VOLTAGE in EOS-1, or by or using the MEMO menu to load a predefined state. 20kV is a good setting to begin with. The voltage is displayed in the information lines of the basic screen, and also in EOS-1. Note that higher voltages will give increased resolution, but also may damage your sample.

9.3.3.8        Check that the objective aperture you desire is in place. One of four apertures (1, 2, 3, or 4) can be chosen. Aperture 1 is the largest and 4 the smallest. Generally 1 will be used for low voltages or high probe currents whereas 4 will be used for high voltages or low probe currents. When you have applied an accelerator voltage and chosen a condenser lens setting, the optimum aperture will be displayed under OPT APERT in EOS-1. If you know in advance what aperture you wish to use, you can select it before applying an accelerator voltage.

9.3.3.9        To change from aperture 4 to 3 for instance, gently pull the aperture selection knob on the SEM column straight out. When the number 3 is visible, rotate the control toward you; it will lock in place. To change from aperture 3 to 4, simply rotate the knob away from you and it will automatically move inward. Remember that the turbo pump is very sensitive to jarring of the SEM, so be careful when changing the aperture!

9.3.3.10  Rotate FILA (filament current) knob clockwise slowly while watching both the screen and the filament current meter. BE CERTAIN YOU ARE LOOKING AT THE FILAMENT CURRENT (FIL BUTTON LIT); IF YOU ARE ACCIDENTALLY LOOKING AT THE EMISSION CURRENT, YOU RISK BURNING THE FILAMENT OUT! You should see the signal on the screen move upward, peak at a maximum near 1.9 A, and then decrease sharply. Center the filament knob on this peak - this is the highest brightness and will give you the best resolution or current-vs.-spot size for lithography. Running above this setting will damage the tip and increase the spot size for the given current into the sample. NEVER INCREASE THE FILAMENT CURRENT ABOVE 2.0A; DOING SO WILL DRASTICALLY SHORTEN THE LIFE OF THE FILAMENT. If the line scan is flat and shows no response to filament current, there are four likely culprits. The secondary electron detector may be off; the Probe Current Detector may be intercepting the beam (PCD in); the beam blanking may be set to ON or EXT; or the gun tilt and shift settings may be far off (see Section 4.4 below). You should also check the SEI brightness and contrast. The Emission Current Meter should read about 40 microamps at 20kV with standard bias selected. If not, report this in Faults.

9.3.4    Aligning the Beam

9.3.4.1        To align the beam straight down the column, press the SHIFT button so that it is dimmed. The tilt/shift knobs will now change the gun tilt. Adjust the PROBE CURRENT knob clockwise; this changes the coarse condenser lens (CL COARSE) setting so as to minimize the probe size. Choose a high setting (e.g. 16). Equivalently, you can also use EOS-1 to set CL COARSE. Adjust the X and Y tilt/shift knobs to maximize the signal on the LH CRT. You may find it easier to find the optimal tilt and shift settings by maximizing the current detected by the PCD. This current is measured by a Keithley picoammeter located on top of the SEM control cabinet.

9.3.4.2        Press the SHIFT button again to enable the shift feature (SHIFT button light will go out). The tilt/shift knobs will now change the gun shift. Adjust the PROBE CURRENT knob counterclockwise until the signal drops near to zero. This increases the probe size and is equivalent to decreasing the CL COARSE setting in EOS-2. Use the X and Y tilt/shift knobs to maximize the current. Continue to reduce the CL COARSE setting and maximize the signal with the tilt/shift knobs until you have done so at CL COARSE of 1.

9.3.4.3       Repeat the adjustment of gun tilt (TILT button lit) at high CL COARSE settings and gun shift (TILT button dimmed) at low CL COARSE settings until the settings remain unchanged as you go from high CL to low CL and back. Check to see that your filament current is still "sitting on top of" the high brightness peak by moving the knob up or down slightly while observing LSP.

9.3.5        Filament Temperature Stabilization

9.3.5.1        It will take about 1 hr. for the filament temperature to stabilize. During this interval, the metal and ceramic parts in the gun heat to equilibrium. The optimal gun tilt and shift settings also may require about one hour to equilibrate.

9.3.5.2       If your process is sensitive to changes in the probe current, the best approach is to adjust the filament current and gun tilt/shift as described above.

9.3.6        Viewing the Sample

9.3.6.1        You may now adjust the working distance according to your requirements. Smaller working distances give better resolution, but if you have a large sample that needs to be tilted, consult the chart on the front of the SEM chamber for the safe working distance for your particular sample. This chart details the X, Y, and Z distances allowable for a given sample size and tilt; the black zone indicates the safe operating regions. For small samples, a working distance of 15 mm is a good operating distance. If the sample is not flush with the top of the holder, focus on the high point, read the working distance on the PNU printout or EOS.

9.3.6.2        On the FUNCTION keyboard, set the SCAN option to picture by pressing the PIC button. There are three PIC settings: a full screen view (PIC in EOS-2), and two smaller screen views (RD2 and RD4). Each time you press the PIC button, you will toggle to one of these three views.

9.3.6.3        Confirm the scan rotation is on and that the tilt correct is set to zero (fully counterclockwise). These controls are located in one of the auxiliary panels in the SEM console. The scan rotation changes the orientation of the image on the screen and is useful for aligning sample axes with the screen axes. The tilt correct compensates for the foreshortening of the image of a tilted sample by intentionally stretching the image along the X-axis. If your sample is not in fact tilted, a non-zero tilt correction setting will lead to incorrect magnification along the tilt axis.

9.3.6.4        Using the FUNCTION keyboard, verify that the FINE focus button is off (not illuminated); this disables the fine focus, and allows for coarse focusing of the sample. (The fine and coarse focus are used to adjust the objective focus.) Adjust focus knob until you can see the sample holder. You can toggle between coarse and fine focus by pressing the FINE focus button. You may have to use the joystick to move the stage and sample holder directly under the beam.

9.3.6.5        Move your sample using the joystick to the area of interest. Optimize focus by watching the LH CRT while adjusting the fine focus. Alternatively, you can use the single line scan mode (press the MODE button) to optimize focus. At low magnifications and coarse focus, adjust the focus knob until the signal exhibits the sharpest peaks possible. At low beam current (high condenser lens settings) it may be easier to use the RH CRT to adjust the focus. This screen shows an averaged image and is quite useful when dealing with a noisy image.

9.3.6.6        You must make final alignment of the beam with the objective apertures. Make certain the SEM is in SEM1 mode; do so by selecting SEM1 under EOS MODE in EOS-2. SEM1 disables the image shift capability of the JEOL 6400. If you remain in SEM mode, any image shift present will interfere with the aperture alignment. In SEM mode, you may set shifts to zero, read on EOS page.

9.3.6.7        Find a small speck of bright material on your sample and focus on it. Make this adjustment at a magnification of 5000X or higher. Press the WOBBLE button and look at the LH CRT (not the RH CRT). The WOBBLE button automatically over- and under-focuses the SEM. Misalignment of the beam with the objective aperture will cause the image to shift from side to side (X) or up and down (Y). Adjust the X and Y aperture centering knobs to minimize motion of the image. The X and Y knobs are on the aperture selection control on the SEM column. The Y adjust is on the end of the control, while the X adjust is comes out of its side. When you have finished, return the JEOL 6400 to full SEM mode (select SEM under SEM MODE).

9.3.6.8        Adjust the X and Y stigmator knobs until you get the most "real" looking image (equivalent to changing STIGMA settings in EOS-1). These knobs control the beam shape, changing it from an oval to a round shape (best image is with a round beam). You will get better results if you make this adjustment at a higher magnification (an increase of 10,000 if possible) than you plan to use for viewing or picture taking. You should make these adjustments while looking at a very small feature (about 100 nm in lateral dimensions). To adjust the position of a feature in the screen while at high magnification, it is useful to use the electrostatic image shift described above. The image position can changed with the X and Y IMAGE SHIFT knobs. To activate the image shift feature, press the POSITION button so that it is dimmed.

9.3.6.9        At high magnifications, adjusting the focus and stigmators may not be sufficient to bring the sample into good focus. In this case, you will need to adjust the PROBE CURRENT; this changes the beam's diameter. You will get greater resolution with a narrower beam. To narrow the beam, rotate the PROBE CURRENT knob clockwise. You should now see more detail, but you will also have more noise visible on the screen (if you take a photo, this noise should not appear on the photo). You will have to compromise between the amount of noise and the detail required. Higher accelerating voltages also give higher resolution but less surface contrast due to the greater penetration of the electron beam. Another compromise is required here.

Note: If the image becomes "fuzzy" after viewing it for long periods of time, the area may be charging with electrons that are not drained away to ground. When this happens, you should move to another location for a few minutes. When you return to the area of interest if it is still not in focus, then the area has been irreversibly contaminated.

9.3.7        Taking a Picture

9.3.7.1        There are 2 screens from which you can view the image and take a picture. The left screen shows a "real-time" image while the right is an FIS (frame integrated storage) image. You can use the FREZ button on the JEOL keyboard (DISPLAY field) to freeze the image on the right screen for taking more than one photo.

9.3.7.2        Press the SLOW button on the FUNCTION keyboard (SCAN field).

9.3.7.3        Optimize the signal contrast and brightness by pressing the WFM button on the JEOL keyboard (DISPLAY field). 5 straight, horizontal lines will appear on the left screen. Use the BRIGHTNESS and CONTRAST control knobs to adjust the scan signal so that it falls within this field during the entire scan. The height of the signal is controlled by the CONTRAST knob, while the overall position of the signal is controlled by the BRIGHTNESS knob. The highest peak of the scan signal should just hit the top line of the grid, while the lowest point should be positioned at the bottom line. Alternatively, you can try using the ACB (automatic contrast and brightness) feature to optimize the signal for picture taking, but the success of this method depends on the sample.

9.3.7.4        Press the WFM button again (light will go out) and the image will return to the left screen.

9.3.7.5        Load the Polapan 400 film into the camera to the right of the viewing screens. To do so, switch the lever to "L" (load), and push in the film until it clicks in place. Then pull on the film casing, which will slide partway out, exposing the film inside the camera.

9.3.7.6        Press either the LEFT or RIGHT button on the JEOL keyboard (PHOTO field) to initiate taking the photo. The LEFT button takes the signal from the real-time image on the left screen, while the RIGHT button will use the signal from the FIS screen. Wait until the scan is complete. To process the film, push the film casing back in, switch the lever to "P" (process) and pull the film out firmly and evenly.

9.3.8        Turning Off the Beam

9.3.8.1        Load the 0kVWD39 state from the MEMO menu. This performs several functions: it returns the SEM to 0 kV without turning the accelerating voltage off; it changes the magnification to 300,000X; it turns the SEI CONT and BRIGHT down; and it changes the working distance to 39mm. There are good reasons for these settings. Leaving ACC VOLTAGE on but at 0 kV leaves some current running through the filament, keeping it clean and preheated. Turning up the magnification reduces heating of the scan coils because at high magnifications, the beam scan area is small. Turning down the SEI CONT and BRIGHT reduces the chance of accidentally burning a spot in the CRTs. Finally, setting the working distance to 39 mm leaves the SEM ready to view a just-loaded sample.

9.3.8.2        Rotate FILA current knob fully counterclockwise.

9.3.8.3        Press SLOW button on JEOL keyboard (SCAN field).

9.3.8.4        Turn off the secondary electron detector by selecting OFF under DETECTOR in EOS-3.

9.3.8.5        Return to the basic screen (press PF1), hide the information bar on both CRTs (use ESC and BREAK), and put the cursor on the left screen. Turn down contrast and brightness of both CRTs. These steps are designed to prevent burn-in of images in either CRT.

9.3.9        Removing the Sample and Pumpdown

9.3.9.1        Set the stage to its home position by typing “h” at the keyboard of the stage control computer located to the right of the SEM; this should automatically set the X and Y at the home or center position (25 mm on the X and 35 mm on the Y travel counter). Manually adjust the tilt and rotation counters to zero.

9.3.9.2        Return the working distance (Z height) to the specimen exchange distance of 39mm; the red light on the front of the airlock exchange panel will come on to indicate that the working distance is correctly set to 39mm. If you have adjusted the fine Z position, return it to 2.0 mm.

9.3.9.3        Hold the glass viewport with the clip facing towards the SEM chamber against the o-ring seal at the airlock entrance. Press the illuminated white button on the airlock chamber and wait for the light to go out. The light is on a timer and should allow enough time for the airlock to evacuate. You should also visually check the o-ring for wear or contamination. If this is not the case, report the problem on FAULTS. Improper evacuation of the airlock can cause the turbo to "dump" when you open the airlock door resulting in down time!

9.3.9.4        When the light goes out, open the airlock door by rotating the airlock knob towards you 1/4 of a turn. The vacuum gauge on top of the SEM will briefly rise to the 10-2 torr range when the door is opened. Slide the door open by pulling the knob to the right. The light in the SEM chamber should come on when you open this door. If the light is not on, then the secondary electron detector is on; turn it off by selecting OFF under detector in EOS-3.

9.3.9.5        Carefully push the sample rod into the chamber and thread the rod into the dovetail holder on stage. Make sure that the rod is completely screwed into the holder before retracting or your sample may fall off into the chamber.

9.3.9.6        Gently retract the rod until the screw is secured in the clip attached to the glass viewport (you will hear a click when this happens).

9.3.9.7        Slide the airlock door closed and turn the knob ¼ turn away from you until the flat area on the knob is parallel to the floor.

9.3.9.8        Holding the rod near the glass viewport, press the white airlock button and wait for the light to go on. Make sure that you hold the rod during the venting of the airlock.

9.3.9.9        Remove loading rod, unscrew the sample holder and place the rod in its holder.

9.4   Electron Beam Lithography with NPGS

The Nanometer Pattern Generation System (NPGS) from Nabity Lithographic Systems is a flexible system for performing electron beam lithography. The purpose of this manual is to provide users with a brief overview of the operation of NPGS v7.6. After reading this, you should still study the following sections of the NPGS manual itself:

Introduction

Microscope Considerations

Designing Patterns

Making a Run Parameter File

Writing Patterns

Section 11 (Sample Preparation and SEM Setup) is also useful reading. If you need to align your patterns with structures already present on your samples, you should read Section 8 (Aligning Patterns). Finally, the program NPGS.exe (only a part of the whole NPGS lithography system), while not required for exposure, adds some flexibility and is described in Section 6. Copies of the NPGS v7.6 manual are available for reading in the Microlab lobby.

9.5        Drawing Patterns with DesignCAD

9.5.1        Before doing an exposure, you must make patterns, which can be interpreted by NPGS. NPGS is designed to accept patterns drawn in DesignCAD, a commercial CAD program. This manual does not attempt to describe in detail how to make drawings in DesignCAD, but does list some of the most useful commands. A copy of the DesignCAD manual is available in the Microlab lobby. It is advised that you study the tutorials in order to familiarize yourself with basic use of DesignCAD before trying to make any patterns.

9.5.2        A directory "pgxxx" will be created for each user, where xxx are the user's initials. You should save you patterns and run files in your directory, and run the NPGS software from it as well. To start DesignCAD from the DOS prompt type "dcxxx" from any directory. This loads a customized version of DesignCAD, which includes several useful macros for use with NPGS. Design CAD will now automatically save your patterns as *.dc2 files, an ASCII file type which can be interpreted by the Nabity software. Your patterns will be saved to your directory by default. You can use many of the drawing features of DesignCAD when making patterns, including lines, polygons, arcs, and cubic splines. See the NPGS manual (available in the Microlab lobby) for details. The most common DesignCAD commands can either be selected from the menus using the mouse, or entered from the keyboard (usually a single stroke).

9.5.3        Some Useful Commands:

9.5.3.1        0/Ins or left mouse button: Sets a new point at the position of the cursor

9.5.3.2        . (period): nearest existing point. Known as a "gravity point."

9.5.3.3        : (colon): Sets a point at absolute coordinates entered by the user.

9.5.3.4        V: connects all existing points in the order they were drawn by a series of straight line segments. Useful for drawing lines; width of lines may be specified by user.

9.5.3.5        ALT F5: Uses a set of points chosen by the user to define a polygonal region. Such regions are treated as filled by NPGS. Known as "PolyFill."

9.5.3.6        Q: Among other things, changes the current color and line width.

9.5.3.7        L: Changes the active layer.

9.5.3.8        ESC: Removes the most recent structure (point, line, etc.) added to the drawing. Pressing ESC a second time will delete the second most recent addition, etc.

9.5.3.9        !: The inverse of ESC.

9.5.3.10     Y: Clears the screen.

9.5.3.11     F10: Saves the current drawing.

9.5.3.12     F9: Retrieves a drawing and overlays it on the current drawing.

9.5.3.13     F8: Exits DesignCAD.

9.5.3.14     Most patterns not requiring curved lines can be drawn using the absolute coordinate (:), gravity point (.), vector (V) and PolyFill (ALT F5) commands. IMPORTANT: When interpreting DesignCAD files, NPGS assumes the drawing units are in microns.

9.5.4        Drawing Layers and Colors

9.5.4.1        NPGS differentiates between the various elements of a pattern based on the drawing layer and color of each element. For all elements in a given layer, you use the program MRF to assign the following exposure parameters: "Origin Offset", "Magnification", "Center-to-Center Distance", "Line Spacing", and "Measured Beam Current". Drawing elements are restricted to layers 1 through 19. Within a given layer, you use MRF to assign a different exposure time/dose to each color you have used. NPGS uses only colors 1 through 16. The same color can be used for different doses in different layers. Layers are exposed sequentially in numerical order.

9.5.4.2        A typical design strategy is to put the most sensitive (i.e. smallest) elements in layer 1, which is to be drawn at high magnification. Larger structures are placed in higher layers, and drawn at lower magnification with higher probe current. Exposure can be paused between each layer, allowing microscope settings (typically the magnification, probe current and/or objective aperture) to be changed.

9.6        Run Files

9.6.1        After you have saved a DesignCAD file, use the MRF program to assign values to the various exposure parameters. The proper syntax is "mrf" or "mrf run_file"; be sure you are in your directory and that copies of the patterns you want are present. If you name an old run file you will be asked if you wish to read the old run file (default is yes) and if you wish to reread the pattern data (default is no). A screen will then appear in which you can select which patterns are to be exposed. You can expose up to 16 distinct patterns in a single run file. Add or remove patterns with the INS and DEL keys. If the pattern list takes up more than one page, the PgUp and PgDn keys will move to previous and following pages.

9.6.2        Once all the patterns have been chosen, enter the various exposure parameters for each. "Ctl PgDn" will call up the run parameter table for the first pattern. This includes the following parameters, for each layer in the pattern:

"Origin Offset": Offsets the origin of the layer during exposure.

"Magnification": Sets the magnification for the layer.

"Center-to-Center": sets the spacing between adjacent points in a line.

"Line Spacing": sets the spacing between adjacent lines in an area.

“Measured Beam Current": Used to calculate exposure time if a dose is specified.

9.6.3        In the first line of the run parameter table you may choose among four options: "w", "p", "c", and "s". "p" pauses the exposure process before any given layer until you tell it to continue. The default "w" causes the layer to be written without a pause, using beam blanking between exposure points. The program will normally pause before the first layer of each pattern in the run file, even if "w" is chosen. This automatic pause can be overridden using the "Pause only for 'p'" option; see the NPGS manual for details.

9.6.4        After the list of layer run parameters, all the colors found in that layer are listed. There are three ways to specify the exposure level for each color: 1) set the exposure time per point; 2) set a line dose in nanocoulombs per centimeter; 3) set an area dose in microcoulombs per square centimeter. The latter two methods are generally more useful; pressing the space bar while the dose line is highlighted toggles between them. When a dose is specified, the exposure time per point is calculated from the measured beam current, the center-to-center spacing and the line spacing (see NPGS manual for details).

9.6.5        If the parameter table for a pattern takes up more than one page, "PgDn" and "PgUp" will move among the pages. After all the parameters for a pattern are specified, subsequent patterns can be reached by pressing "Ctl PgDn". "Ctl PgUp" will return you to the previous pattern.

9.6.6        After specifying the run parameters for all patterns, save the run file by pressing "Ctl Home". Pressing "Ctl End" will exit MRF without saving. Run files are saved with the suffix ".rf6".

9.7        Exposing Run Files

9.7.1        Once the run parameters have been set using MRF, the program PG is used to expose the pattern. The syntax is "pg run_file" where run_file is a .rf6 file created by MRF. Be sure you are in your directory and that a copy of the run file and all patterns in it are present. A screen will appear which lists the pattern name, layer number, magnification and measured beam current specified in the run file. Several options are now possible. Pressing the space bar will expose the pattern; "!" will skip the current layer and move to the next in the run file; "Esc" will skip the current pattern; and "Enter" will cause PG to terminate without exposing the pattern. PG can be terminated during an exposure by pressing "Ctl Break".

9.7.2        You can also use the NPGS.exe program to perform exposures. It gives added flexibility, including the option to perform DOS commands between patterns. See the NPGS manual for details. If you wish to align your exposure with previously existing structures, you will need to make use of the alignment program AL. See the NPGS manual for details on the creation of alignment patterns and the use of AL.

9.8        System Operation for Electron Beam Lithography

System operation for electron beam lithography parallels that for general use, with some additional steps and precautions. This section is not stand-alone. Liberal reference will be made to Sections 2 though 4 above. See also the NPGS v7.6 manual, Section 11 (Sample Preparation and SEM Setup). For a general introduction to electron beam lithography, see Introduction to Microlithography, L.F. Thompson et al., Eds. (1983).

9.8.1        Sample Preparation

9.8.1.1        Samples must be clean and free from contaminants. Samples should be coated with an electron beam resist such as PMMA or P(MMA-MAA). If an electron microscope resolution standard is mounted along with the sample you may use the standard for focusing and astigmatism adjustment. No particles of anything, silver, graphite or other will be used. A scratch made on the surface is fine, so is the edge of the chip. Silver particles in methanol were once used for focusing. This procedure resulted in migration of the silver particles with downtime and maintenance needed for cleanup.

9.8.1.2        It is useful to put two focusing marks on the sample, one on each side of the exposure area. This allows an estimate of the correct fine focus setting at the exposure area without risking exposure of the sample. Also, since the resist is sensitive to electrons, the sample should be mounted on the stage so that you know roughly its stage coordinates. This allows you to move the field of view off th