Chapter 9.1

Karl Suss BA6 Bond Aligner

(ksba6)

 

1.0        Title

Karl Suss BA6 Bond Aligner

2.0        Purpose

The Karl Suss MA/BA6 can be used in  two  modes:  mask  align (MA6)  for lithography and bond align (BA6) for wafer-to-wafer alignment followed by clamping or prebonding. This chapter describes the bond alignment capability of  the  tool (BA6).  For mask align mode refer to the Chapter 4.17. The Karl Suss BA6 is  a  bottom  side  wafer-wafer  bond aligner  with alignment accuracies down to 1 micron or better. Only trained and approved users may use this bond aligner. The system includes a number of unique features including:

     Bottom side alignment using bottom viewing optics,  CCD imaging  with  frame  grabber and TV monitor. This allows registration of features on the backside of one wafer  to the topside of the other wafer (BACKSIDE ALIGNMENT MARKS OR THROUGH-WAFER ETCH WINDOWS ARE NEEDED FOR ALIGNMENT).

     Automatic computer  control  with  LCD  status  and  user prompting  and  keypad  entry.  Up to 100 programs may be stored with 7 on line at any one time. Each  program  may be edited for key parameters pertinent to the bond mode.

     Anodic bonding (transparent glass substrates to  silicon) and  silicon  fusion  bonding programs are available with options for slide/fixture load/unload. With  the  current machine   configuration   LOADING  IS  ALWAYS  BY  SLIDE. Unloading is  by  slide  only  for  silicon  fusion  pre-bonding. Unload by fixture when performing clamping following by anodic bond in the SB6 The seven programs on-line are summarized as follows:

Available Programs:

1.      Anodic bonding/Prox: Load slide - Unload fixture. The only parameters that needs to be entered is alignment gap and WEC type. Proximity WEC is usually selected.

2.      Silicon fusion: Load slide - Unload slide. Used for Si-Si pre-bonding, typically before the wafers are annealed in the furnace for fusion bonding. Since the wafers are actually pressed against each other, in addition to alignment gap, the pre-vacuum, vacuum and purge times have to be entered. This is similar to vacuum contact program in MA6 mode.

Programs NOT used:

Silicon fusion: Load fixture - Unload fixture

Not used.

Silicon fusion: Load slide - unload fixture

The only parameter that needs to be entered is alignment gap.

Silicon fusion: Load fixture - Unload slide

Not used.

Anodic bonding/Prox: Load fixture - Unload fixture

Not used.

Anodic bonding/Prox: Load fixture - Unload slide

Not used. Laser pre-bond option is necessary which is not available with the current system.

     Wafer sizes can be either 4 inch or 6 inch.  Change  over from  4  to 6 inch and vice versa may be made in a minute or two. For other substrate sizes special chucks  can  be ordered.

     Fast or slow scanning of  mask/substrate  with  a  memory toggle  to  quickly  view  between  2  locations (as when checking alignment key and some other known critical pattern). Aids alignment using one objective.

3.0    BA6 Setup

Before starting, verify that the machine is  in  the  correct  idle state:

     System power is ON and the  machine  is  in  the  **MA6** mode.

     TV monitor is off.

     Lamp power is on with supply readout at 900  in  constant intensity mode 1.

     Mask and wafer trays are on  stage  with  dummy  mask  no wafer is on the chuck.

     Pressure = 5-6 bar; Nitrogen = 1-2 bar.

     When you enable machine look  out  for   any  problem reports.

Start-up Procedure

(1)         While machine is in MA6 mode, remove  mask  holder  by pressing "CHANGE MASK" button. Place mask on the table to the left of stage.

(2)         Turn machine off  by turning the  green  knob  on  the display panel.

(3)         Turn machine back to ON. Hold knob in ON position for a  couple of seconds until "KARL SUSS precision instruments" is displayed. If  this  is  not  done  the  breakers  at  the  back  of the machine might trip and need to be reset.

(4)         Press LOAD when  prompted.  Machine  will  go  through starting up routine.

(5)         Select **BA6** machine configuration using Y arrows to toggle between BA6/MA6 modes. Press ENTER.

(6)         Machine will ask user to push lamp housing all the way to the back. Push it back manually (very carefully and slowly) until a clicking sound is heard and  the  lamp house locks in place.

(7)         Press F1 to bring up microscope  UP/DOWN  menu.  Press enter to move microscope up.

(8)         Remove the mask adapter frame by unscrewing  the  nuts, which  hold  the  frame  onto  the stage. The frame is quite heavy and very  expensive  piece  of  equipment. EXERT CAUTION WHEN HANDLING IT. Place the frame on the shelf below the machine making sure it is  stable  and secured in place.

(9)         Select the appropriate bond  adapter  frame  from  the slots  on  the shelf below the machine. The frames are bulky (silver color) and easy to identify. The smaller  (4")  frame  is  used  for  silicon fusion pre-bonding only. Anodic bonding of both 4 and 6" wafers  use  the 6" frame.

(10)      Place the frame onto the stage. Make sure  it  secured properly onto the stage. Also make sure the clamps are in the up position. If they are  not  you  can  do  it manually.

(11)      Select the proper upper  chuck  for  your  application from  the  chucks stored in the storage cabinet behind and to the left of the  machine.  For  silicon  fusion *pre-bonding* i.e. load slide - unload slide you would use the Si fusion upper chuck. For other  applications you  would  use the 4" transport fixture, which is usually found sitting on the transport  arm  of  the  SB6 bonder.  Place chuck carefully taking care not to damage the clamps and/or spacers. IMPORTANT: THE FLAT  ON THE CHUCK WITH THE NOTCH SHOULD BE TOWARDS YOU.

(12)     Replace bottom chuck with the proper  chuck  for  your application. For silicon fusion pre-bond you would use the 4" vacuum bond chuck with rubber seal.  For  other applications  using  the  transport fixture use the 4" bond chuck. Store away the MA6 vacuum chuck  you  just replaced.

4.0    Viewing, Loading, Saving and Editing a Program

The BA6 allows 7 different program recipes (described earlier) to be in online memory at one time.  Any one of these programs can be immediately chosen by simply pushing  "Select  Program" pad until the appropriate program appears in the display. Up to 100 programs (numbered 0-99) may be stored in archived memory. These programs must be loaded before they can be executed.

To see what programs are stored:

(1)         Press "Edit Program". Pad flashes.

(2)         Hit Y (up/down arrow) keys. Each program #, bonding  mode &  alignment  gap is displayed as you scroll through each program. You can scroll up or down through  the  list  by pressing the Y-up or Y-down key.

(3)         To exit looking at programs, press  flashing  "Edit  Program" key.

To load an already stored program:

(1)         Push "Edit Program". Pad flashes.

(2)         Press X-lateral key to scroll between: 1 "Save Prgm  to"; 2  "Load Prgm from"; 3 "Delete Prgm #"; 4 "Exit Prgm editor without change". Scroll to 2, Load Prgm from.

(3)         Press Y key until desired program number appears  in  LCD display.

(4)         Press flashing "Edit Program" key to load stored  program into online memory.

Example: To load program 9

(1)    Press "Edit Program".

(2)    Press X-lateral key once to scroll to "Load Prgm from".

(3)    Press Y-up/down key 9 times to scroll to recipe 9.

(4)    Press "Edit Program" again to load recipe and  exit  edit mode.

To change a stored program:

(1)         Load desired program as described previously.

(2)         Press "Edit Parameter" key, light flashes.

(3)         Press X-lateral key to  scroll  through  parameters.  For soft  or  hard contact modes parameters are: Expose Time, Alignment Gap and Wedge Error Compensation (WEC) Mode.

Note:    For vacuum mode there  are  additional  parameters. Select  parameter  or  parameters to be changed. Saving a program also saves the objective position.

(4)         Press Y-up/down key to change parameter to desired value.

(5)         Press "Edit  Parameter"  key  again.  Change  appears  on display  and  light stops flashing. Now you still need to save this program if you want to  use  it  again  in  the future.

To save it as another program:

(1)         Press "Edit Program" key. Light flashes.

(2)         "Save Prgm to" is displayed.

(3)         Press Y-up/down key to select program # to save to.

Note:    You  can  save  a  program on top of another program, but first make sure that program is not needed. If  you  wish to  save  to  a  particular  program # AND that recipe is empty simply press "Edit Program".

(4)         Press "Edit Program" key again to execute save.

To Delete program (to free up 1 of 99 slots  for  a  new  program):

(1)    Press "Edit Program" key. Light flashes.

(2)    Press X-lateral key twice to scroll to option 3,  "Delete Prgm #"

(3)    Press Y-up/down key until  program  you  wish  to  delete appears.

(4)    Press flashing "Edit Program" key again.  Flashing  stops and program is deleted.

 

To Exit the "Edit Program" mode without making a change:

After entering the "Edit Program" mode by pressing "Edit  Program" key:

(1)    Press X-lateral key until option 4, "Exit Prgm Editor..." is displayed.

(2)    Press "Edit Program" key again. Light stops flashing  and edit mode is exited.

 

To make a change without loading a new program:

Example: To change alignment gap for the next substrate only:

(1)    Press "Edit Parameter" key. "Alignment gap" is  displayed as the 1st parameter to change.

(2)    Press Y-up/down key to increase/decrease gap.

(3)    Press flashing "Edit Parameter" key again. New align. Gap should be displayed.

5.0    Wafer Loading

There are 2 bottom wafer bond chucks, both  4  inch,  one  for fusion  pre-bonding  (this  one  has an inflatable rubber seal ring) and the other for all other  bonding  applications.  The vacuum grooves on the chuck must be completely covered by your substrate. Make sure you have the right  chuck  installed  for your  application. If a different wafer chuck is desired, push straight up on chuck and remove. Place desired chuck into circular  opening  and  align white line on chuck to pin on tray. Always handle wafer chuck by  the  rim,  do  NOT  touch  chuck directly.

CAUTION:   These chucks are  made  by  precision  machining  and should be handled with care. Do not drop, scratch or otherwise damage them. Store unused chuck in proper location.

(1)         Adjust wafer rotation knob to 0 (so that rotation  finger is  centered)  and set X & Y micrometers for the stage to 10.

(2)         Push "Load" key. The  display  will  ask  you  to  insert proper  bonding  frame.   Since  you have already done so press "Enter".

(3)         The display will then ask to  insert  the  proper  bottom chuck. Since you should already have installed the proper chuck press "Enter".

(4)         Place your substrate with the side to be bonded down making sure all vacuum grooves are covered. The wafer should sit between the 4 positioning pins with steel balls  with the flat towards you.

(5)         Press "Enter" and slowly push tray all the way  into  the machine.

(6)         Push "Enter". The wafer  is  positioned  onto  the  upper chuck  by vacuum and control lever of the proximity flags is activated.

(7)         For silicon fusion bonding programs you  should  now  use the  BSA  microscope  to find the alignment marks on your wafer. Once you have found the marks and  adjusted  focus the  image should be grabbed. BSA set-up and operation is described in later sections. For anodic bonding  applications,  the  program  assumes that you have a transparent bottom wafer and does not give the frame grabbing  option at  this  stage.  In  this case you would load the second wafer and then use BSA to grab image.

(8)         Pull out slide and load 2nd wafer with  the  side  to  be bonded up.

(9)         Press "Enter"  and slowly push tray all the way into  the machine.

(10)     Press "Enter". The  second  wafer  is  brought  into  its uppermost  position  and  WEC is performed. After WEC the proximity flags are retracted and wafer is brought to its align position.

6.0    BSA Microscope Set-up and Wafer Alignment

The bottom microscope (BSA) is used to align the  backside  of the bottom wafer to the front side of the top in silicon fusion bonding mode. It accomplishes this by using a CCD camera, which screen grabs,  or  stores an image of the wafer PRIOR to putting your second wafer in. The microscope is then locked into place while  your second wafer is loaded. The second wafer must have backside alignment marks or etched windows. The  second  wafer is  then  viewed through the quartz chuck by the bottom microscope and alignment is done by moving the  WAFER  ONLY.   Both the  mask's  stored  image  and  the  real time wafer image is viewed on the TV monitor.

(1)         Verify machine idle mode and start up machine  as  previously described.

(2)         Select desired bonding program and parameter settings  as described above.

Note:    This step is very  important  and  should  be  done before loading your wafers.

(3)         Load first wafer as previously described.

(4)         Press "BSA/IR" illumination switch (image can be seen  on monitor).

Note:    Top microscope is never used in BA6 mode.

(5)         Activate bottom microscope movement controls by  pressing "BSA  Microscope" pad (Key illuminates). The X-Y keys and "Fast" keys may now be used to move bottom microscope for scanning the wafer.

(6)         There are 3 keypads "Left, Both and Right"  which  determine  mode  of movement; "Left" -for left objective movement, "Both" - for both objective  movement  etc.  Select desired  mode  first  before using X-Y arrow keys to move bottom microscope.

(7)         Find the alignment marks on the wafer.

(8)         Press "Top/Bottom"  key  to  activate  wafer  focus  (Key should illuminate) if it is not already on.

(9)         Set  the  3-position  field  switch  to  desired   value (Left = left  objective,  center = split  field,  right = right objective). Split field mode is the recommended mode.

(10)      Adjust "Upper Substrate" left and right  focus  knobs  to get a sharp image on monitor.

(11)      Adjust left and right illumination knobs to adjust lighting.

(12)      Bottom microscope  position  may  be  read  out  on  LCD, Xl = left objective x position, Xr = right objective x position, Yl = left objective  Y  position  etc.   Rotation  is accomplished  by  moving  one  objective  relative to the other.

(13)      Once you have moved to and focused your  alignment  marks press "Grab image" to capture image.

Note:    This must be done every time you load a new wafer.

The mask and microscope are now locked down and  may  not be moved. To cancel "Grab Image" and restart, press "Grab Image" again, make your adjustment and press "Grab Image" a third time.

(14)      Load second wafer as described in the previous section.

(15)      Once the second wafer is loaded "Top/Bottom"  key  should be  in bottom mode (Light NOT lit). If light is on, press key.

Note:    If you are using the anodic  bonding  program  then you will grab image now and not in the previous steps. At this stage you would:

(1)         Select "Top/Bottom" key so that is lit  i.e.  top substrate is selected.

(2)         Move BSA microscope objectives  and  find  alignment marks  on  your  top (Si) wafer (looking through the glass wafer) as described above. Grab  image  of  Si wafer alignment marks. Upon completion of this step, the objectives of the BSA microscope  are  automatically focused to the plane of the glass wafer.

(16)      Rotate "Lower substrate" focus  and  illumination  knobs. (Use  both left and right knobs if you are in split field or either knob for the appropriate single objective.)

(17)      All movement now is done by moving the wafer, using  wafer X,  Y  and R knobs. If too much resistance is felt during wafer movement, or alignment gap needs to be changed  for any  reason, unload and change program parameter and load again. Align appropriate alignment marks  on  the  bottom wafer  to  alignment  marks  on captured top wafer image. Verify  appropriate  program  and   alignment   gap   are selected, if not you will have to unload, correct program and load again.

(18)      For silicon fusion pre-bonding applications you will  now press  "Pre-bond".   The  wafers will be pressed together with full vacuum and the lowered down as a stack  to  the slide.  For  anodic  bonding  or fusion bonding involving unload by fixture jump to step (20).

(19)      Pull substrate slide out and unload wafer stack and press "Enter".

(20)      For anodic bonding or fusion bonding with unload by  fixture,  press  "Perform  Clamping".  The  2 wafers will be securely clamped in the transport fixture.

(21)     Unload fixture and confirm with enter.

7.0    Machine Shutdown

When you are finished bonding all of your substrates:

(1)         Remove your last substrate by sliding wafer tray straight out  and removing wafer stack. If unloading fixture, confirm with enter.

(2)         Push tray all the way into the BA6.

(3)         Press flashing "Enter"

(4)         Turn the machine off.

(5)         Remove the bond adapter frame and  put  it  back  in  its place below the machine.

(6)         Put the mask adapter frame back on the stage.