Chapter
9.2
(ksbonder)
1.0
Title
Karl Suss (Suss MicroTec) Bonder User’s Manual
2.0
Purpose
This document contains information regarding to
KSBonder equipment capabilities, operation instructions, processing parameters,
and basic theoretical discussions of various bonding process.
3.0
Scope
3.1
Overview
The KSBonder is a semiautomatic, computer-controlled, stand-alone
substrate bonder equipped with a vacuum/pressure chamber and a loading arm. The
machine processes aligned and unaligned wafers, substrates and chips. The
alignment accuracy of this tool is listed as being 3 µm (3 σ).
All bonding pair alignment is done on the KSBA6 tool (Chapter 9.1), the substrate stacks are mechanically clamped using the transport fixture, and then transported and bonded in the SB6 chamber.
3.2
Equipment Capabilities
► Vacuum
base pressure: 5.0E-5
Torr.
► Temperature
controller limit: 500ºC
► Bonding
Force up to: 60 psi
(750 lbs for 4’’ wafers)
3.3
Materials Allowed (It is very
important to have similar thermal expansion characteristics of the bonding
pair)
►
VLSI
materials: Silicon, PolySilicon, Silicon Nitride, Silicon Oxide
►
Photo-resist
(temperature no higher than 150ºC)
►
Metals:
Aluminum, Gold, Gold-tin alloys, Lead-tin alloys, Indium, Copper
►
Glass:
borofloat, Pyrex, 7740, Hoyag SD1, or any borosilicate glass with thermal characteristics
matched to silicon
►
Quartz
3.4
Configurations and Components
3.4.1
Windows
NT Based Control and Data Acquisition System
3.4.1.1
Each
user will have his/her personal NT account and SB6 software interface account
after passing the qualification exam.
3.4.1.2
Each
user will have his/her personal file directory for recipe and data storage.
3.4.2
The
Motorized Z drive allows different bonding sequences.
3.4.3
The
Compressed air and N2 pressure controllers need to be turned on
prior to the process.
4.0
Applicable Documents
Link to anodic bonding research report done by Ning Chen.
5.0
Definitions and Process Terminology
For anodic
bonding, the working theory states that applying a large voltage potential
across the glass-wafer complex generates an electric field that drives Na+ ions
into the glass wafer away from the interface region. Thus an Na+ depletion zone
is formed leaving oxygen molecules at the interface. Oxygen molecules are
then diffused into silicon to form a layer of amorphous SiO2.
The setup
on KSbonder is shown in Figure 1. There are two ways to orientate the bonding
pairs.
5.1
With
Glass – Silicon Configuration (Negative Polarity)


5.2
With
Silicon – Glass Configuration (Positive Polarity)


There are some differences between the two setup configurations. Negative polarity configuration with Pyrex on top, as shown in Figure 1, has a longer bonding time and better bond quality than the configuration shown in Figure 2 (Pyrex on bottom, positive polarity). For more information see the linked research report from Chapter 4.
6.0
Safety
Use common sense and general caution when approaching, and working at the machine. Watch out for potential “pinch points,” e.g. pneumatically actuated door. There is no toxic gas used at KSbonder.
7.0
Statistical/Process Data
8.0
Available Process, Gases, Process
Notes
8.1
Available Tooling
8.1.1
UNI TOOL: 4’’ and 6’’ substrate capacity anodic and
thermocompression bonding. The Uni-tool uses one central electrode for anodic
bonding applications. This can be contrasted with the Anodic-tool, which has
two concentric sets of electrodes. Because of its versatility, the Uni-tool is
the default installed tooling at ksbonder.
8.1.2
TC TOOL: 4’’ substrate capactity thermal compression
bonding. Note this tool rarely used since Uni-tool performs same function.
8.1.3
ANODIC TOOL: 4’’ substrate capacity anodic bonding
tool. Has two concentric sets of electrodes, purportedly to enhance bonding
rate and bond quality.
8.2 Recommended
recipe for 4’’ anodic bonding:
►
Voltage: 1500 V
►
Polarity: Negative
►
Pressure: Purge/Atmospheric
►
Temperature: 450ºC
These recommended parameters from results of a series of studies. For
more details, please contact Matt Wasilik.
9.0
Equipment Operations
9.1
Startup
Procedures
9.1.1
Enable
ksbonder..
9.1.2
Pull
out, and turn on the compressed air and N2 gas until gauge reaches the marked
pressure set point.
9.1.3
Login
to the Windows NT environment.
9.1.4
Click
on SB6 Start icon on the desktop to start SB6 main menu.
9.1.5
Login
from the navigation bar with your own user name and password.
9.2
Downloading
Recipe
9.2.1
Click
the CONTROL CENTER icon located in the navigation bar.
9.2.2
Click
on Download Recipe button in the control center window.
9.2.3
Browse
through the recipes in the current user directory and select the appropriate
recipe for the process. (It is recommended to verify the recipe prior to load.
This is described in Section IV).
9.3
Loading
Bonding Pairs
9.3.1
After
align the bonding pair on SB6 aligner, position the fixture apparatus on the
loading arm.
9.3.2
Click Load
on the control center window.
9.3.3
Wait
until the process chamber door opens, then push the load arm into the chamber.
9.3.4
Click Move
on control center window and wait until the stage moves up and then release
the fixture from loading arm.
9.3.5
Pull
out the loading arm and click Finish in control center window.
9.3.6
Click Start
to run the recipe
9.4
Monitoring
the Process
9.4.1
Click
on SB6 from the task bar.
9.4.2
In
this window, many parameters are monitored and displayed: chuck temperature,
chamber pressure, voltage, current, and etc.
9.4.3
Many
parameters can also be monitored in continuously updated plots (located on the
bottom of the screen).
9.4.4
The
bonding process is completed after reaching the end point setting. This setting
will be discussed in detail in the recipe Section 9.7
9.4.5
A data
file is automatically generated during each execution of a recipe. This file
can be reviewed from Data icon on navigation bar. The file can be
also saved on to a floppy disk by click on Copy icon on the screen shown
below. The data file is only available after the process is finished.
9.5
Unload
and Shutdown Procedure
9.5.1
Wait
for the chamber to cool down to 200ºC.
9.5.2
Click Unload
to initiate the unloading process.
9.5.3
Once
chamber door opens, push the load arm into the chamber.
9.5.4
Click Move
to set the bonding pair fixture on the load arm.
9.5.5
Move
load arm out of the chamber when prompt to do so.
9.5.6
Click Finish
to close the chamber and allow machine to return to an idle state.
Temperature would automatically drop down to 100ºC.
9.5.7
Logout
the SB6.
9.5.8
Shutdown
the software.
9.5.9
Logout
Windows NT environment.
9.5.10
Shut off the compressed air and N2 valves to prevent an “over
pressure” error condition in the chamber.
9.6
Aborting
a Process
For
whatever the reason should the user feel the need to abort a process, do not
use the Abort command on the control center window. The acceptable
alternative is to Pause the process and Skip through each
of the subsequent steps.
The navigation bar has RECIPE option that allows user to create,
edit existing recipes. All recipes for the current user are stored in d:/sb6/users/username/current/ directory. If the user wishes to copy other
user’s recipe, please do so in the windows explorer window and paste it in to
the current user recipe directory.
Recipes consist of a number of sequential actions executed at specific conditions (i.e. pressure, temperature, membrane force, and voltage). Each recipe step consists of several parts. To create/load a recipe, click on RECIPE icon located on the navigation bar.
To edit existing recipe, first highlight the recipe and then click Edit.
This would bring out a separate recipe screen. In this screen, each of the
steps can be edited. In a typical recipe, there are eight steps.
9.7.1
Bringing
tool in contact with bonding pairs.
9.7.2
Removing
clamps for the bonding pairs.
9.7.3
Removing
spacers.
9.7.4
Bringing
the inner circle electrodes in contact with bonding pair to initiate the
tacking process.
9.7.5
Brief
pausing.
9.7.6
Actual
bonding process.
9.7.7
Cooling
down to 200ºC.
9.7.8
Unloading
wafers and bring machine to default state.
Each of the steps can be edited by clicking on the parameter (i.e. pressure, temperature, and etc.). For example, if the user wishes to change temperature setting for a specific step, by double clicking on the temperature window, it will bring out a new temperature edit window (shown below). In this window, the use can: set the temperature; giving a tolerance; ramp up time. After changing the parameters, click OK to save. If the user wishes to apply this setting to other setups, just click on Next All. Note that tolerance settings in the recipe parameters are critical. Be sure to have the appropriate values set! Ignoring the tolerance settings can lead to unexpected process behavior.
The end point setting is available in the actual bonding step. There are three different ways to define the end point:
a.
Percent
of max current.
b.
,
the slop of current curve.
c. Area under the curve (Q, total
charge).
9.8
Changing
Tooling
The default tooling configuration at ksbonder is Uni-tool. That is to
say, when approaching the machine in a default state, the Uni-tool will be
installed. Users are allowed to change
to the anodic tooling at ksbonder only when properly trained! Note that
when the anodic process is finished, the Uni-tool must be reinstalled by
default. Anodic tooling install directions follow:
9.8.1
Ensure that the chamber is at atmospheric pressure,
and that the tooling is at room temperature. In the ksbonder software enter the
CONFIG menu. Under the” Current Tool name” option, select “anodic tool”. Then
select “replace tool” option. Shut the software down and turn off the computer.
Press EMO button to turn off all power to the ksbonder.
9.8.2
Two persons are generally required for the following
procedure. Use the lift mechanism to carefully raise the enclosure. Be careful as to not bump or scrape it on
any of the components just under the enclosure. There is a delicate PCB mounted
in the right hand rear corner of the chamber lid surface that can be easily
damaged if care not taken.
9.8.3
Once the enclosure is lifted, use hex key to loosen
and remove the bolts, one at a time, that secure the Uni-tool to the chamber. Be careful not to drop the bolts!
Disconnect the pneumatic Pa (uni tool) line, and lines 8.1 and 8.2.
Disconnect the E1 and E2 electrode connections. Disconnect the JBMEM
and JBC electronic connectors. Once connections are removed the Uni-tool
may be gently lifted up and out of the chamber (this generally requires two
persons – Uni-tool weighs roughly ~25 lbs). Be
careful not to drop the tool or bump, scrape it against any other components.
9.8.4
Remove the struts that the anodic tool rests upon
and re-secure to Uni-tool. Ensure o-ring is placed on anodic tool before
insertion into chamber. Wipe all sealing surfaces clean with IPA and tech
wipes. Gently place the anodic tool into the chamber. Be careful! Do not drop the tool into the chamber! Note there are
alignment marks on tool and chamber to ensure correct orientation. Connect the E1
and E2 electrode connections to the anodic-tool. Install all of the
bolts to the anodic tool head. They should be tightened in sequence via a “star
shape”. Use hex key to tighten ¼ turn past being finger tight. Do not over tighten!!!
9.8.5
Carefully recheck all the connections and bolts.
When confident everything okay, gently and slowly lower the enclosure until it
is secure. Again, it is important to use
caution – components can be easily broken if care is not taken. Turn
equipment power back on and reboot computer. Log in to software and verify that
anodic tool is seen by software in the CONFIG menu previously described.
Finally, insert the red anodic tooling sign on the enclosure cover.
9.8.6
When finished processing with the anodic tool,
replace with default Uni-tool. (inverse of instructions detailed above). Do not
leave the anodic tool in the chamber.
10.0
Troubleshooting Guidelines
Known
Bugs: If the
compress air and N2 gases are not turned off when machine is in idle
state, it will result in over-pressure problem. Therefore, please make sure that all gasses are turned off after
the process is complete.
Software
Problems: If the
software refuses to cooperate, press CTRL+ALT+DEL to close all programs. Log
out from Windows NT and log in again. Re-start the program. If the problem
persists, try to do a hard reboot of the PC system.
Things
To Avoid:
(1)
Avoid
using abort command while running recipes. The recipe will try to reach the
specified parameters fro the step before aborting. Pausing and stepping through
each step seems to be more effective.
(2) Ensure that tolerances in recipe
settings are correctly set. Do not ignore tolerances, as unexpected equipment
behavior can result.
11.0
Figures & Schematics
12.0
Appendix
Rev. 00 – 3/03 – N. Chen
Rev. 01 – 6/06 – M. Wasilik